Taking out the trash: Guidelines for agentic single-trace BLI sensorgram processing

Author: Benjamin Osborn, PhD and the Gator Bio team


Overconfident BLI hit assignment impacts AI training

In June 2026 the Boltz team published a candid look at how the binder-design field reports its results [1]. They noted two common points where real-time kinetics results get misread: ambiguous sensorgrams getting logged as confirmed binders, and affinities getting calculated from curve fits that don’t match the data, which makes their affinities look tighter than they are. The Boltz team proposed a nomenclature fix that can help to clarify these results: calling ambiguous cases screening hits, but reserving the term “binder” for conclusive data. In the process of validating questionable hits, they used Gator® Bio Biolayer Interferometry (BLI) instruments to invert ligand and analyte, which highlighted how kinetics can change based on how assays are designed and formatted.

Experimental hit rates for AI-designed binders vary widely by project [2-4], but even in the best cases many candidates fail binding or developability criteria, so wet-lab screening remains a decisive bottleneck. And when models are trained on datasets that label candidates as binders without adequate justification, measurement errors and artifacts become overconfident training labels that bias the model’s behavior.


[BACK TO TOP]

A rules-based approach for BLI hit triage

Existing high-throughput kinetics-analysis workflows show that preprocessing, curve fitting, residual review, and automated reporting can be standardized at scale [5-7]. This article complements that literature by addressing a narrower problem: how to interpret a first-pass BLI screen that contains only one sensorgram per candidate at a single analyte concentration. The goal is not to generate precise KD values from those traces, but to perform a rapid quality audit and assign each candidate meaningful flags and labels that inform the next steps. Instead, the goal is to sort each trace into one of the following categories:

  • Hits: Clean candidates that pass the screening criteria
  • Questionable hits: Candidates that pass the screen, but have one or more suspicious attributes
  • Negative: Traces that do not achieve the hit threshold
  • Excluded: Unusable traces that defy BLI conventions

In addition to categorizing each trace, these rules will also inform downstream assays by searching for key criteria (e.g. biphasic behavior, off-rates, and bulk shifts). These in turn decide what the next experiment should be, whether it’s a classical titration for a clean hit, steady-state assays for fast-on/fast-off binders, or inverted assays for binders with unusual binding. While these pseudocode blocks are pressure-tested against real Gator instrument data, this is only meant to be a starting point for developing an AI-based BLI workflow for single-trace hit sorting.

To help both humans and AIs understand the logic, there’s a section of prose explaining the rationale for these screening rules and a pseudocode section that outlines the logic clearly. Users can then query an LLM by attaching the gator data file as a .zip and using the prompt below to perform the analysis.


[BACK TO TOP]

Assumptions:

The first step, not covered in this article, is running a good kinetics assay. Each of these is standard practice an experienced BLI user knows by rote, but an automated analysis strategy can only work if they’re in place. Do not mistake the brevity of this section for unimportance.

  • Nonspecific binding has been screened out. You’ve confirmed the analyte doesn’t stick to the sensor surface on its own, independent of the ligand.
  • The ligand is the molecule being screened; analyte is consistent. This is typical for BLI first-pass screening. Often, ligands come from cell-free media and each is screened with only one biosensor at one (often unknown) concentration. Analytes are usually at 0.1-1 µM concentration depending on the quality of starting ligands.
  • Ligand and analyte masses are known, though they’re only used for plausibility flags.
  • Binders are ideally monovalent, such as VHHs, scFvs, Fabs, or designed miniproteins: Full-length antibodies often show biphasic association or dissociation from avidity; the logic tolerates this but flags it.
  • The load step is present and readable, so the logic can confirm a ligand actually captured onto each sensor.
  • Reference subtraction is done beforehand. We recommend having a single trace with an unrelated ligand interacting with the analyte to account for the “jumps” from bulk shift as you enter and exit those wells. An unreferenced (or improperly referenced) plate may run into complications from baseline drift or bulk shift. We include baseline drift check as part of the fitting equation in case this is not performed.

The workflow at a glance:

This workflow is broken into two section, labeled B1 and B2 (because A will be used for steps that precede this in a later article). B1 cleans and fits the data, then decides what candidates are unusable and which ones are usable, but did not clear the hit floor. Several other flags are also fired in B1 that label the data for later sorting. B2 then takes the information that B1 has given and assigns both hit classes and priority for follow-up experiments. A flowchart of the workflow is included below in Figure 1.

Flowchart of the single-trace screening logic. B1 checks usability and the hit floor, and B2 sorts hits by class and priority.

Figure 1: The workflow at a glance. B1 checks for whether a trace contains meaningful data, while B2 assigns priorities and next steps (colors match later figures).


[BACK TO TOP]

Resolving bulk shift

The first thing we need to ask when examining a sensorgram is where association and dissociation actually start. The sensorgram often shows jumps going into the association and dissociation steps, caused by changes in refractive index or by binding that happened during the unread time while the sensor moved into a new well. These jumps can also appear in a reference trace, which then carries them into every trace that used that reference.

We only have one trace’s information to work with to resolve this, so we measure the size of the jump on entry to the association well and on exit from it. A pure refractive index change goes up on the way in and comes back down by the same amount on the way out, so if the two jumps are opposite in direction and within 0.9 to 1.1x of each other, we treat the jump as bulk shift and start the fit from that offset. We then measure the hit against that corrected starting point, so a bulk shift can’t trigger a “hit” without any real binding.

When the jumps don’t match, the direction of the mismatch tells us something. A larger entry jump suggests binding happened during the dead time before the first read, which is exactly what a very fast-on, slow-off binder looks like (JUMP_IN_EXCESS). A larger exit jump suggests signal was lost during the dead time on the way out, as with a very fast off-rate (JUMP_OUT_EXCESS). In either case, we use the smaller jump as the bulk shift estimate, since the larger one likely includes binding gained or lost while the instrument wasn’t looking. If both jumps go the same direction, it isn’t bulk shift at all; that points to a step artifact or a reference problem, and we flag it JUMP_SAME_SIGN, leaving the trace uncorrected, though we still try to see if it has useful signal.


[BACK TO TOP]

Preventing low-effort fitter failures

Every trace gets the same fitting routine, which checks whether the binding curve behaves like 1:1 binding (or close to it). Association fit is a pseudo-first-order exponential (a global Langmuir fit needs multiple analyte concentrations; this is the single-trace application of the same model), plus a baseline drift term for ligands that slowly leave the sensor or plates without reference subtraction. Dissociation is fit as a single exponential decay. Both fits start from the bulk-corrected values described above.

To keep fit failures rare, the first attempt is seeded from the data itself (e.g. the time to reach two-thirds of the rise). If it fails, we retry from several other starting points, then fall back to a model without the drift term. This catches the vast majority of BLI binders and includes samples with what appears to be a linear association phase (they just have a small koff)

Some real binders don’t follow a single exponential; full-length antibodies often show a fast phase and a slow phase from avidity. When a single-exponential fit converges but exceeds the deviation limit, we try a two-phase fit with clear rules: both phases must move in the same direction, their rates must differ by at least 3x, and each must carry at least 5% of the signal and clear the noise threshold. Passing traces are flagged ASSOC_BIPHASIC or DISSOC_BIPHASIC, so the titration step won’t assume a 1:1 model.

If every fit still fails, we try a flat line. If even that can’t describe the data within the deviation limit described below, the trace is erratic and excluded as ASSOC_FIT_FAILED or DISSOC_FIT_FAILED.

If a flat line does fit the association, the trace is flagged ASSOC_UNRESOLVED. This will flag when a binder is so fast that its binding happens during dead time between wells. This logic provides a failsafe to protect very tight binders, but it requires us to confirm that binding happened in that dead time. The jumps from the bulk shift detection phase tell them apart: a larger entry jump (JUMP_IN_EXCESS) followed by a much smaller out-jump and accompanied by a slow or unmeasurable dissociation means the trace is kept on the strength of its off-rate alone. It may be a tight binder, but with a single trace, we can’t know enough about it to be sure. If the jumps match or there’s no slow dissociation, the trace is classified BELOW_HIT_FLOOR.

If a flat line fits the dissociation, the signal simply isn’t coming off within the window, which we treat as a slow off-rate rather than a failure (covered in “Reading the dissociation phase” below).


[BACK TO TOP]

What is a significant deviation from a binding trace?

Real binding has a characteristic shape: during association the signal rises and bends toward a plateau, and during dissociation it decays smoothly, one direction at a time. That regularity gives us something to test against. A real hit, even a noisy one, should stay within three times its noise of the fitted curve, and a trace that wanders, spikes, or refuses to settle into a binding shape is excluded. The same test applies to dissociation.

We anchor this deviation limit to noise so it’s neither too harsh on noisy traces nor too loose on quiet ones. Gator Bio instruments are specified at ≤ 4 pm RMS noise at 5 Hz, and most traces run quieter than that, so we use the spec as a floor: the tightest limit is 3 × 0.004 nm = 0.012 nm. Noisier traces (for example, crude lysate) get leeway up to 0.010 nm RMS, for a limit of 0.030 nm. In the code, RESID_CAP = 3.0 is the multiplier, and the noise is clamped between 0.004 and 0.010 nm.

This residual-to-noise check is our goodness-of-fit test. We avoid R², which can look acceptable on a poor fit when the binding signal is large; scaling the misfit to noise is closer in spirit to a reduced χ². Since our model is a deliberate approximation, a formal χ² analysis isn’t worth the compute, and this check handles triage well enough.

Here’s one that fails. Its association passes cleanly, but then the sensor enters a well with a pipetting error. Its deviation from the decay fit is 9.5 times noise, a two-phase fit can’t rescue it, and it ends below the second baseline, earning a DISSOC_RESID fail. The cause could be an instrument fault, a bubble, the plate wall too close to the sensor tip, or insufficient well volume. The test doesn’t diagnose which; it just excludes the trace.

BLI sensorgram of Assay 4, Channel 5. The association fits well, but the dissociation deviates 9.5x noise and is excluded as unusable.

Figure 2: example of a trace that fails the significant deviation test

These fail codes also help an agent spot spatial problems on the plate: if ASSOC_RESID or DISSOC_RESID failures cluster in one region, that points to an instrument or liquid-handling issue, and those wells can be scheduled for a re-run.


[BACK TO TOP]

Reading the dissociation phase

At screening concentrations (as high as 1 µM), an analyte can sit at 100x or more above the KD of quality binders, which can produce unusually shaped curves that are hard to tell apart from artifacts. With only one trace per candidate, we can’t rely on association for resolvable kinetics, so we turn to dissociation, usually a simple single exponential decay.

Our check is the time to reach 90% dissociation, calculated from the fitted off-rate. If the signal is 90% gone within about a second, the kinetics can’t be measured. We use 1 second as a practical default for 5 Hz acquisition, since it leaves only five data points to define a decay. That corresponds to a koff of about 2.3 s⁻¹ and an average residence time of 0.43 s, beyond the instrument’s koff spec.

We sort each dissociation into one of four states:

  • FAST: 90% of the signal is gone within about a second, including signal lost before the first read (JUMP_OUT_EXCESS).
  • NOT_RESOLVED: The off-rate is slow, but the step was too short to measure it, so the titration should run longer. This triggers when less than 5% comes off within the window, including flat lines or upward drift during dissociation.
  • RESOLVED: everything in between, with enough decay to support the fitted off-rate.
  • NO_SIGNAL: association never produced enough real binding to dissociate, regardless of bulk shifts.

NOT_RESOLVED is where the tightest binders on a plate usually land. Their off-rate isn’t measurable yet, but B2 still treats them as “slow-off”, so they stay eligible for high priority.

For an example of fast-on/fast-off kinetics, consider the two panels below. They come from the same small-molecule plate, and both pass the deviation test cleanly. The trace on the left is a hit we can characterize: association curves toward a plateau, and both rates are slow enough to read. The trace on the right rises and falls almost vertically, with a t90_off under 1 second. The binding may be real, but under these acquisition conditions, it can’t support a reliable off-rate estimate. It’s labeled FAST, and B2 sorts it into HIT_FAST_OFF at low priority. In a protein screen, an interaction that fast is usually weak; small-molecule campaigns might weigh it differently.

Two BLI small-molecule traces. Left: a binder with a readable off-rate. Right: a fast-off trace with an off-rate too fast to resolve.

Figure 3: Left: an example of a fast-on/fast-off interaction that is measurable. Right: a trace that is unmeasurable due to an off rate that is too fast. This trace is moved later to low priority for a steady-state screen. Both traces come from a Gator BLI small molecule experiments using SMAP (Small Molecule Analysis Probe) biosensors on a GatorPlus instrument.


[BACK TO TOP]

What’s a hit?

We want to set a “hit floor” high enough that baseline drift won’t trip it by accident, but low enough that even a weaker ligand could trigger it. For most campaigns, a simple 0.1 nm (or higher) hit floor makes sense. It’s 10x above the maximum effective noise used by this workflow, and it’s more than any amount of baseline drift typically produced during a screening workflow’s association step (less than 180 seconds). For campaigns featuring smaller analytes or unusual ligands (like aptamers), a more permissive hit floor of 10x above noise (e.g. 10x measured noise per trace, clamped between 0.004-0.01 nm noise to control for high/low noise samples) would remain rigorous while still allowing for small analytes that may never reach 0.1 nm shifts even at saturation. An alternative to make a more stringent can be based on ligand/analyte size, and comparing overall analyte signal to load signal as a ratio.

When deciding whether a trace achieves “hit” status, we look at the average of the last 5 seconds of data (measured from the bulk-corrected starting point). This lets us avoid noise on the last data point tripping the “hit” detector.

Not every real binder clears the hit floor. A tight hit with a fast on-rate can reach its plateau during the dead time before the first association read, so at screening concentration its association looks flat or falls short of the floor. Two signals give it away: the entry jump is larger than the exit jump (JUMP_IN_EXCESS), because binding was added before we started reading, and the dissociation is slow or NOT_RESOLVED. When both are present, we keep the trace on the strength of its off-rate alone. B2 classifies these as HIT_FAST_ON_SLOW_OFF at medium priority; they could be the best binders on the plate, but a single trace can’t confirm it. If you’re screening at lower concentrations, this check may not be necessary.


[BACK TO TOP]

B1: the first batch of pseudocode

There are a few pieces in the pseudocode that we didn’t talk about in the prose:

  • Step and setup information carried with each trace: downstream analysis shouldn’t depend on the original assay.asy file always being in a particular place.
  • HIGH_RMS, DISSOC_RISING, BASELINE_DRIFT, DISSOC_BELOW_BASELINE: quality flags that annotate a suspect trace without excluding it, giving more tools for troubleshooting.
  • LOW_LOAD: a sensor that captured some ligand, but less than 0.1 nm or too little to clear the hit floor even at full binding (based on masses of ligand and analyte). The trace is kept, but a “non-binder” call on it may reflect poor loading rather than poor binding. NO_LOAD, by contrast, means the load was within noise and the trace is excluded.
  • The hit floor is signed, so it also catches downward signals. Binding adds mass, so a hit is normally a rise above the floor, but unusual ligands (like aptamers) can produce real negative signals. We keep these and flag them NEGATIVE_BINDER, and B2 classifies them as HIT_QUESTIONABLE at low priority. A downward signal is more often the result of drift, a reference problem, or the analyte pulling something off the sensor, so the flag also helps trace plate positions when troubleshooting referencing errors.

The values below in the CONSTANTS section are practical defaults for the assay format and acquisition rate described here, not universal BLI cutoffs. Each is tied to a stated physical or statistical rationale (above), and users are encouraged to adjust them to adapt to their assay workflow.


[BACK TO TOP]

B2: Sort the survivors

Everything that reaches this point is a hit, and a secondary experiment is needed to confirm a “binder.” B1 has already set aside the unusable traces and the non-hits, so B2 only decides two things for each survivor: what kind of hit is it, and what’s next? Many screens produce more hits than can be tested, so B2 also assigns a priority. Classes are checked in this order, and the first match wins:

ClassConditionNext experimentPriority
HIT_QUESTIONABLEReason: NEGATIVE, SUB_BASELINE, or KINETIC_CONFLICTOrientation flipLow
HIT_FAST_OFFt90 under 1 sSteady-stateLow
HIT_SLOW_ONLinear association, consistent slow offTitrationMedium
HIT_FAST_ON_SLOW_OFFBuried association, JUMP_IN_EXCESS, readable off-rateTitrationMedium
HIT_MODERATE_OFFCurved association, t90 between 1 and 30 sTitrationMedium
HIT_CLEANCurved association, t90 ≥ 30 s or NOT_RESOLVEDTitrationHigh

The low-priority classes come first on purpose, so a trace with a problem can’t be labeled a good hit by an earlier check.

HIT_QUESTIONABLE traces point to the sample or the measurement rather than a well-behaved binder. KINETIC_CONFLICT is the new reason: since kobs = kon·C + koff, the off-rate can never exceed the observed association rate (we allow 1.25x for measurement error), so an association slow enough to watch paired with a dissociation too fast to read is physically inconsistent. These route to an orientation flip, which is costly enough to keep them at low priority.

HIT_FAST_OFF traces lose 90% of their signal in under a second. The near-vertical, square-wave shape reflects real fast-on/fast-off binding, but usually a weak binder, so they go to a steady-state measurement at low priority.

HIT_SLOW_ON traces rise in a straight line: an expected outcome at 100 nM; a very slow on-rate at 1 µM. A rise that hasn’t completed one time constant means kobs is below roughly 1 / (association time), so koff must be too. Traces that pass go to titration at medium priority; those that fail are a KINETIC_CONFLICT. Aggregation can also look like a slow, consistent linear rise, so the titration has the final word.

HIT_FAST_ON_SLOW_OFF hits could be tighter than any clean hit on the plate, but because the on-rate is “buried” in dead time, a single trace can’t prove it. Placing them at medium priority rather than high is a judgment call on our part: they’re worth a titration, just not ahead of hits we can already characterize.

HIT_CLEAN is the straightforward case: a curved association and a slow dissociation (a t90_off of 30 seconds or more, or NOT_RESOLVED) point to a likely good binder. HIT_MODERATE_OFF has the same characteristics, but reaches 90% dissociation in 1 to 30 seconds, which is measurable but points to a weaker binder, so it goes to titration at medium priority.

A few items in the B2 pseudocode aren’t covered elsewhere in this article.

  • RMAX_IMPLAUSIBLE flags a hit whose signal exceeds what the loaded ligand should support; it never changes the class, since BLI mass ratios are approximate, but it can assist agentic processing. plate_inputs_ok is a switch that the Rmax check reads, asking whether a whole-plate sanity check confirmed the mass and valency inputs.
  • ASSOC_NOT_SATURATED fires when an association never reached a plateau, a note to the titration step to use a longer association or a higher concentration.

These connectors are here because this stage is one piece of a larger pipeline, and it’s cheaper to leave them in place now than to retrofit them later.


[BACK TO TOP]

Applying the sorter

Figure 4 is an experimental set with 48 different antibody ligands against the same analyte run on a Gator Plus instrument. I ran this data through B1 then B2 and asked the LLM to apply a colored border based on the outcome. Most read as non-binders (grey), exactly what a yes/no screen should produce. The survivors sort cleanly into the HIT_CLEAN (dark green) and HIT_MODERATE_OFF (light green) classes. We also see a few HIT_SLOW_ON (purple) on this plate, which makes sense given that the analyte was run at 100 nM. These may be slower-on binders, but their dissociation is slow and consistent, so they’re worth a titration. A messier plate, or one with faster kinetics, might also populate the steady-state and flip classes too.

A keen eye might point out a sample like Assay 4, channel 5: why is it still getting classified as a high priority binder when it just barely clears the “hit” bar? This is because we’re using a uniform 0.1 nm hit threshold, and because its t90_off is about 60 seconds, past the 30 second cutoff. In all respects, it meets our HIT_CLEAN parameters and the pseudocode is behaving as intended. However, going into an assay like this, I might set a slightly higher hit threshold given the size of the analyte. That said, be wary of setting a dynamic threshold that relies on the association height of the other hits on the plate (e.g. cutting all hits below 25% of the median of all hits). Simple logical rules of this nature can backfire and omit perfectly good binders when all hits are around the same intensity, or when there are few hits on the plate.

Grid of 48 BLI antibody screen traces, color-coded by B1 and B2 outcome, with 11 high-priority hits.

Figure 4: A 48-trace BLI high throughput single-trace screen featuring 11 high priority hits. No referencing was used for these samples.

Figure 5 is a 60-antibody screen against a 100 nM analyte concentration, with 300-second association and dissociation steps and four empty sensors averaged as the reference. Here, the sorter prioritizes each trace as expected. Some traces that look like fast-offs, such as A4Ch1, reach 90% dissociation at 34 seconds, just past our 30-second cutoff, so they’re classified as high priority alongside the other binders in green. Eight candidates survive, all high priority, and the remaining 52 fall below the hit floor as clean negatives. Assay 3 Channel 1 and Assay 8 Channel 4 both show the slowest off-rates (t90 of 250 and 350 seconds, respectively), making them strong candidates for follow-up.

Grid of 60 BLI antibody screen traces against a 100 nM target, with 8 high-priority hits and 4 averaged reference sensors.

Figure 5: a 60-antibody high throughput single-trace screen with 8 high priority hits identified. Samples A8Ch5-A8Ch8 were averaged together as reference traces.

To get an example of clear biphasic behavior, we ran it against a dataset that features Fc receptors as ligand and full-length IgG as analyte. While the assay was originally a titration experiment (each assay its own titration with one Ab tested in 1-3 and another in 4-6 against three different Fc receptors), the instructions to the LLM included a statement to use only channel 1 (the top concentration of each titration series: 30 µM for Assays 1, 2, 4, and 5; 10 µM for Assays 3 and 6) and to use Channel 8 from each assay used as reference.

As we explain in the assumptions section, most screening campaigns won’t see this kind of fast-on, biphasic binding. However, the logic correctly assigned Assays 1 and 4 as HIT_CLEAN at high priority and 2 and 5 as HIT_MODERATE_OFF based on the dissociation rate. Assay 6 falls below the hit floor, a clear negative. Assay 3 is the interesting one. It clears the hit floor, but its fitted off-rate exceeds its observed association rate, which points to a problem with the sample or the measurement. B2 classifies it as HIT_QUESTIONABLE (KINETIC_CONFLICT) at low priority, since the check can’t tell which.

Six BLI traces of two antibodies against Fc receptors, showing clean, moderate-off, questionable and below-floor outcomes.

Figure 6: two antibodies tested against Fc receptors. Only top concentrations are considered, with reference channels applied per assay (not shown). Assays 1-3 are Ab1 against three different Fc receptors; Assays 4-6 are Ab2. This was an added test to confirm functional sorting logic in B1/B2 for biphasic binders.

Finally, we have an example of some uglier data, Figure 7. This is a kinetic screen of 17 antibodies captured on the sensor, run against a single target protein. We used only the top concentration (100 nM) and subtracted a sensor that captured no antibody as the reference. Four antibodies (Ab13, Ab15, Ab16, and Ab17) are HIT_CLEAN: they curve toward a plateau and barely dissociate over five minutes. Five more (Ab1 through Ab5) clear the hit floor but rise in a straight line without curving, and none of them dissociate measurably in the window (NOT_RESOLVED). That pairing is self-consistent, so B2 classifies them as HIT_SLOW_ON at medium priority.

One lesson from this dataset was how much the reference matters: the first choice of reference had a small step at the start of dissociation, which the fitter read as a fast off-rate and flagged a kinetic conflict on two good binders. A cleaner reference (A5Ch7, an irrelevant antibody sensor that captured almost nothing), made both conflicts disappear.

Grid of BLI antibody kinetic screen traces at 100 nM, showing clean, slow-on, negative and excluded outcomes.

Figure 7: Example dataset with 17 different binders. This result required using a different trace than what is labeled as “blank” (A5Ch5) to obtain a usable reference. After using A5Ch7 for reference, all traces resolved cleanly.


[BACK TO TOP]

Some caveats:

Automated analysis rules and flags can cut down the workload for a reviewer, allowing them to quickly train an AI model with this starting foundation. However, this workflow cannot replace a trained kinetics expert.

Some things to bear in mind (which we’ll cover in more depth in a follow-up article):

  • Referencing: A reference sensor loaded with an irrelevant ligand usually matches the sample surface better than a blank biosensor. A loaded sensor seeing zero analyte is also not a sufficient reference. It may account for baseline drift, but it doesn’t see the bulk shift jumps as you enter the high analyte concentration. If a consistent set of spikes are seen across candidates, try double-referencing using a ligand that showed no binding and see if that improves your fits.
  • Rebinding: on a densely loaded sensor, analyte that comes off can rebind a neighboring ligand before it escapes. Both slow the apparent off-rate and make a trace appear to have a slower or biphasic off rate. This means that the koff from the initial screen may not align with the final koff of a titration performed at a concentration closer to the KD.
  • DISSOC_RISING: flags traces whose signal climbs during dissociation, which can mask a real off-rate. If it appears across many traces that share the same analyte or buffer, the buffer itself is the likely cause (for example, a bad lot of BSA), and the fix is a fresh buffer or a better-matched reference rather than re-running the candidates.

[BACK TO TOP]

A step toward automated rigor

None of the individual rules here are unique. A bench scientist makes most of these calls by eye without thinking about them. However, at screening scale nobody has time to make those calls by hand across hundreds of candidates, so the work here was less about inventing new science than about writing down what a careful analyst already knows.

This piece is in the middle of the workflow, focusing on the steps of garbage collection and hit sorting. Future articles will examine:

  • Pre-flight quality checks for BLI data and how to set up a high throughput screen with the greatest chances of success
  • What a follow-up assay looks like for each condition (steady-state vs kinetics titration) and how to know you’ve generated good data.

In the meantime, feed the logic above into your favorite LLM, pull one of your own screens (or download this one, corresponding with the 48 trace panel above), and see where it agrees with you and where it doesn’t. We’d love to hear how it performed: reply to our post on LinkedIn and let us know.

To hear more about how Gator Bio is powering AI research and hit screening, check out our recent recordings from Closing the Loop SF 2026, a gathering on the AI-native lab.


[BACK TO TOP]

Works Cited

[1] Boltz team. De novo binder design model evaluation. Boltz; 2026 [cited 2026 Oct 5]. Available from: https://boltz.com/experimental-protocol

[2] Erasmus, M.F., Bedinger, D., Hopkins, E. et al. A blinded, prospective benchmark of in silico antibody discovery anchored to experimental affinity and developability. Nat Biotechnol (2026). https://doi.org/10.1038/s41587-026-03238-6

[3] Bennett NR, Coventry B, Goreshnik I, et al. Improving de novo protein binder design with deep learning. Nature Communications. 2023;14:2625. doi:10.1038/s41467-023-38328-5.

[4] Dreyer FA, Schneider C, Kovaltsuk A, et al. Computational design of therapeutic antibodies with improved developability: efficient traversal of binder landscapes and rescue of escape mutations. mAbs. 2025;17(1):2511220. doi:10.1080/19420862.2025.2511220.

[5] Srinivasan B, Flórez Weidinger JD, Zhai X, Lemercier G, Ikeda T, Brewer M, Zhang B, Heyse S, Wingfield J, Steigele S. High-throughput mechanistic screening of non-equilibrium inhibitors by a fully automated data analysis pipeline in early drug-discovery. SLAS Discov. 2022 Dec;27(8):460-470. doi:10.1016/j.slasd.2022.09.003. Epub 2022 Sep 23. PMID: 36156314.

[6] Dahl G, Steigele S, Hillertz P, Tigerström A, Egnéus A, Mehrle A, Ginkel M, Edfeldt F, Holdgate G, O’Connell N, Kappler B, Brodte A, Rawlins PB, Davies G, Westberg EL, Folmer RH, Heyse S. Unified Software Solution for Efficient SPR Data Analysis in Drug Research. SLAS Discov. 2017 Feb;22(2):203-209. doi:10.1177/1087057116675316. Epub 2016 Oct 28. PMID: 27789754; PMCID: PMC5302086.

[7] Li K, Huntwork RHC, Horn GQ, Alam SM, Tomaras GD, Dennison SM. TitrationAnalysis: a tool for high throughput binding kinetics data analysis for multiple label-free platforms. Gates Open Res. 2024 Jun 28;7:107. doi:10.12688/gatesopenres.14743.2. PMID: 38009106; PMCID: PMC10667272.

Blog

Beyond the Snapshot: How Gator Bio BLI Complements Structural Biology

Beyond the Snapshot: How Gator Bio BLI Complements Structural Biology

Author: Benjamin Osborn, PhD


Introduction

Biolayer interferometry (BLI) is a powerful addition to the structural biologist’s toolkit, complementing snapshot techniques like X-ray crystallography and cryo-EM with real-time binding data. BLI’s real-time measurements also don’t require expensive isotopic or atomic labels that drive techniques like NMR and HDX-MS, making it an ideal tool for difficult-to-produce proteins or for studying many similar proteins in succession. However, BLI is not a single-molecule technique – it requires an interaction partner to characterize these protein-protein dynamics. Ten published studies below used Gator BLI alongside the traditional techniques. The examples below start with BLI’s most established structural use, mapping which residues carry an interaction, and finish with its newest. Section 4 covers two papers in which BLI was the first wet-lab measurement a computationally designed binder was subjected to, including the validation set behind BindCraft.

For a more thorough treatment of the methods used, consult the original publications.

The ten papers at a glance

Ref
System Measured
Biosensor
HIV-1 Integrase CTD / Importin 7
Amine Reactive
CCHFV GP38 / survivor antibodies
ProA
J2 IgG / double-stranded RNA
Protein A
ACE2-Fc / SARS-CoV-2 Spike trimer
Mouse Fc
N3-1 antibody / Omicron RBD
Human Fc Gen II; Strep-Tactin XT
Frizzled4 nanodiscs / Norrin, DKK1, DEP
SA; SMAP
Tspan12 nanodiscs / Norrin
SA
Neurotensin receptor 1 / NT(8-13)
Ni-NTA
PD-1, PD-L1, IFNAR2 / BindCraft binders
Protein A
ChuA / hemoglobin and designed binders
Ni-NTA

1 – Mapping interaction surfaces and binding residues

Studying binding interfaces is one of the most common applications of Gator BLI. Often, this takes the form of epitope binning, where researchers run two binders against the same target in sequence and see whether the first blocks the second. However, as crystal structures and in silico approaches grow more sophisticated, mutagenic screens are commonly used to build a map of which residues are critical to the interaction. Protein-protein pairs are the most common, but Gator BLI can be used to study interactions with other biomolecules as well, such as DNA/RNA [see 1-3].

[1-1] – HIV-1 Integrase must reach the nucleus to integrate viral DNA into the host genome, and Importin 7 had long been implicated as a carrier, with two arginine/lysine-rich motifs in Integrase’s C-terminal domain identified as the likely binding site. Researchers at the Hebrew University of Jerusalem characterized the interface directly, combining HDX-MS and crosslinking MS to map contact regions, AlphaFold3 to model the complex, and Gator BLI to supply affinity data. Integrase CTD was immobilized on amine-reactive biosensors via EDC/NHS coupling, and Importin 7 was titrated as analyte across a broad concentration range, with affinities determined by steady-state analysis. Wild-type Integrase CTD bound Importin 7 with a KD of 17.9 nM, while a mutant carrying two substitutions in the second nuclear localization signal (NLS) motif (RRKAK → RAAAK) bound poorly at 299 nM, confirming that the NLS is key to high-affinity binding (See Figure 1).

Figure 1, comprising Figure 5A and 5B from the above publication, showing the steady-state binding of HIV-1 integrase (WT and RAAAK mutant) with Importin 7. Adapted from Bana J, Yariv A, Oppenheim T, et al, FEBS Open Bio. 2026. doi:10.1002/2211-5463.70294, used under CC BY 4.0.

[1-2] – Crimean-Congo hemorrhagic fever (CCHF) is a tick-borne viral disease with a case fatality rate reported up to 40%, and the WHO has flagged CCHFV as a public health risk lacking approved vaccines or therapeutics. An earlier study identified 13G8, a mouse monoclonal antibody targeting the viral glycoprotein GP38 that protects mice from lethal viral challenge. This is notable because GP38 is not a neutralization target, leaving binding data as the primary in vitro readout. These researchers isolated anti-GP38 antibodies from a human CCHF survivor and used a BLI competition assay to determine which antigenic sites they engaged; three of seven competed directly with 13G8, and crystal structures confirmed that the survivor-derived CC5-17 shares that epitope while approaching it offset by 22° (Figs 2D, 3A). To test how much sequence variation the shared site could tolerate, the team used Gator BLI to screen point mutants (Fig 3E), introducing residues from the related Aigai virus into CCHFV GP38 to abolish binding and reciprocally introducing CCHFV residues into Aigai GP38 to restore it. Most single substitutions barely affected 13G8 binding, suggesting several changes would have to occur simultaneously for viral escape, while a single G296K substitution significantly reduced CC5-17 affinity, consistent with the steric clash predicted by its shifted binding angle. In terms of disease treatment, the human-derived CC5-17 bound GP38 far more tightly than 13G8 yet conferred lower protection in mice, which the authors attribute in part to its lower association rate.

Figure 2, from Supplementary Figure 4 in the above publication, showing binding curves for 13G8 mouse-derived antibody (panels a-e) and CC5-17 patient-derived antibody (panels f-i) binding to viral GP38 from various strains. Adapted from Durie IA, Tehrani ZR, Karaaslan E, et al, Nature Communications. 2022;13:7298. doi:10.1038/s41467-022-34923-0, used under CC BY 4.0.

[1-3] – Not every useful binding pair is protein-protein. J2 is the gold-standard antibody for detecting double-stranded RNA in cells, spotting viral replication intermediates, and screening mRNA therapeutics for inflammatory contaminants. Despite three decades of use, nobody had defined what it binds. Researchers at NIDDK used BLI to perform the bulk of that characterization effort. Running J2 IgG as ligand on Gator Bio Protein A biosensors with nucleic acids as analytes, they showed J2 binds dsRNA robustly while dsDNA, ssRNA, ssDNA, and RNA-DNA hybrids gave no appreciable signal, which means that J2 doesn’t detect cellular R-loops (Fig 1C, 1D). Before turning to mutagenesis they ran a series of dsRNA lengths, which identified the practical binding threshold at roughly 14 bp, well below the ~40 bp previously estimated by atomic force microscopy (see Figure 3, below). Based on a dsRNA/J2 co-crystal structure, the team mutated several CDR residues and used BLI to measure the effect. Two tyrosine residues proved essential, with Y50A and Y52A sharply reducing binding and Y50F and Y52F barely registering a change, pointing to van der Waals contacts from the aromatic rings in the minor groove of the dsRNA, rather than the phenolic hydroxyls. They also tried swapping in basic residues at N101 and S33 on the theory that more positive charge should help bind a phosphate backbone. Both compromised binding rather than improving it. Finally, the team checked whether AlphaFold3 could reproduce the complex. It predicted the J2 Fab and its CDR loops accurately, landing within about 1 Å of the crystal structure, but it placed the dsRNA nowhere near the true epitope, which BLI had already established. The authors attribute the miss to how few antibody-nucleic acid interfaces exist in the PDB for AlphaFold3 to train on.

Figure 3, from Figure 1I in the above publication, showing binding curves for different lengths of dsRNA, and 1J, bottom right, showing steady-state binding curves for each. Adapted from Bou-Nader C, Juma KM, Bothra A, et al, Nature Communications. 2026;17:635. doi:10.1038/s41467-025-67414-z, used under CC BY 4.0.


[BACK TO TOP]

2 – Tracking binding across variants and species

Beyond locating a binding site, Gator BLI can track how that site changes over time as a species mutates, or how binding affinity changes between different species. BLI’s plate-based format allows you to run many binders against many targets in rapid succession without having to prep the samples multiple times. The papers below cover viruses evading host antibodies and viruses acquiring the ability to infect new hosts.

The same set of experiments is run in industrial drug development. Before a therapeutic antibody goes into toxicology studies, you have to show that it binds the target in the model species with an affinity close to human. Otherwise you cannot read much into the tox results. You can do all of that on one plate. You run the antibody against the human, cynomolgus, mouse and rat orthologs, and you use the same biosensor chemistry for all four.

[2-1] – SARS-CoV-2 spilling over into animal hosts had been a concern since the start of the pandemic, and mice are a particular worry as a possible host. Mouse ACE2 (mACE2) doesn’t bind ancestral Spike well at all. Researchers at EPFL wanted to know whether the mutations that were driving immune escape in humans were simultaneously expanding host range. They captured dimeric ACE2-Fc on Gator Mouse Fc biosensors and titrated trimeric Spike as the analyte. An initial screen across cat, dog, and mink showed that each of these other species bound wild-type Spike protein. In contrast, mouse ACE2 showed almost no binding in the original Wuhan strain but later variants of concern bound far better (Fig S1). Full titrations showed that wild-type Spike barely bound to mACE2 even at 300 nM, while Alpha, Beta, and Gamma came in between 25 and 45 nM, and Omicron was around 2-3 nM (See Figure 4). Fig S10 features a mutation screen that introduced point mutants N501Y, K417N, or E484K on their own into the WT Spike background and found that single mutants weren’t enough to confer mACE2 binding, but a dual-mutant N501Y/E484K did.

Figure 4, from Figure 1 in the above publication, showing binding curves for various SARS-CoV-2 Spike proteins for ACE2 from mouse and human, alongside the differences in binding residues among several species. Adapted from Ni D, Turelli P, Beckert B, et al, PLOS Pathogens. 2023;19(4):e1011206. doi:10.1371/journal.ppat.1011206, used under CC BY 4.0.

[2-2] – In addition to increasing the affinity of Spike for ACE2 in mice and humans, Omicron also dodged N3-1, a potent antibody isolated from a first-wave COVID-19 patient that had neutralized every variant of concern until then. Only one of Omicron variant BA.1’s fifteen receptor binding domain mutations even contacts the antibody, and substituting that residue back didn’t restore binding (measured using a human Fc Gen II biosensor). Despite carrying fewer mutations than later variants, BA.1 showed a much faster off-rate. This is because a serine-to-proline substitution at position 373 inverts a nearby loop, covering the hydrophobic pocket the antibody reaches into without altering the pocket itself. As a result, the Omicron spikes hold their receptor binding domains down far more often than the ancestral strain, so the arrangement N3-1 needs is rarely exposed. With Strep-Tactin XT biosensors, the team screened every possible substitution at the clashing residue, running the mutants straight from clarified expression supernatants (with no purification step), and several improved affinity, but no amount of engineering of the site of interest overcame the larger conformational rearrangement. Strep-Tactin XT is exclusive to Gator Bio. Direct capture from cell-free and crude expression removes the purification step, and it may explain why three 2026 AI binder-design preprints used Gator Bio platforms to run their binding assays: BoltzGen, BoltzProt-1, and Candido et al. on protein language models.

Figure 5, from Figure 4A in the above publication, showing patient-derived N3-1 binding to wild-type (WHU1) versus Omicron SARS-CoV-2 strains, illustrating how Omicron escapes antibody binding. Adapted from Goike J, Hsieh CL, Horton AP, et al, Communications Biology. 2023;6:1250. doi:10.1038/s42003-023-05649-6, used under CC BY 4.0.


[BACK TO TOP]

3 – Membrane proteins and lipid environments

Where previous sections dealt with soluble proteins, membrane proteins are notorious for being difficult to characterize, and they’re conformationally coupled to the system they’re in, such as nanodiscs [3-1, 3-2] or detergent micelles [3-3]. Membrane proteins are also expensive to prepare and problematic to purify, meaning low yields and a product that is often too fragile to push through a fluidic system. Gator BLI needs 40 µL per well, so a difficult prep goes further. XT Optics gives 3x signal to noise on top of that, which is what turns a low-shift nanodisc measurement into useful results. Together they let you hold a membrane protein in a pseudo-native conformation and still resolve analyte binding against it.

[3-1] – The team at the Weis lab at Stanford developed a way to ensure completed complexes are present in the same nanodisc with a split GFP tether. Quantitative Western blotting confirmed about two Frizzled4 per nanodisc for the homodimer and one Frizzled4 plus one LRP6 for the heterodimer. To show functionality, they used BLI. Biotinylated nanodiscs were immobilized on streptavidin biosensors, and Norrin bound the Frizzled4 homodimer at 0.46 nM while DKK1 bound the Frizzled4–LRP6 heterodimer at 5.6 nM. Then, to measure the kinetics of Dishevelled-2 DEP domain against nanodisc-immobilized Frizzled4-LRP6 heterodimers, they used Gator Bio XT Optics-enabled SMAP biosensors, capped at a 10 nm shift to avoid crowding artifacts from packing nanodiscs too densely. They ran repeated association-dissociation cycles on a single sensor, with a ligand-free first cycle followed by walking DEP from 10 nM to 32 µM. To account for any drift or non-specific DEP binding, they ran a matched empty-disc reference sensor against the same samples concurrently on a separate tip. DEP bound at roughly 200 nM with or without Norrin present, showing that heterodimerization doesn’t change how Frizzled4 recruits Dishevelled. This work predates SA XT. For biotinylated nanodisc capture today we would recommend SA XT, which pairs the same streptavidin chemistry with the XT Optics sensitivity gain that makes a low-shift titration like the DEP series easier to resolve.

Figure 6, from Supplementary Figure 4a and 4b in the above publication, depicting Norrin binding to biosensor-immobilized lipid nanodiscs. Adapted from Bruguera ES, Mahoney JP, Weis WI. Journal of Biological Chemistry. 2022;298(4):101628. doi:10.1016/j.jbc.2022.101628, used under CC BY 4.0.

[3-2] – Norrin activates the β-catenin pathway like Wnt does, but it needs a co-receptor, the tetraspanin Tspan12. As the complex forms, do Norrin and Tspan12 interact with each other? Researchers in the Weis lab at Stanford purified Tspan12, reconstituting it into biotinylated lipid nanodiscs, and immobilized those on streptavidin biosensors, and titrated Norrin as the analyte. Norrin bound at 10.4 nM, and the interaction was stronger in the nanodisc than in a GDN detergent micelle. They then modeled the complex with AlphaFold and confirmed the putative interaction site with mutagenesis. Charge reversals on Norrin alone reduced affinity by up to two orders of magnitude, but compensation on the paired binding site (Norrin K102E/R121E with Tspan12 E196K/S199K) restored binding, confirming the interaction site precisely.

Figure 7, from Figure 1 in the above publication, depicting that the Norrin-Tspan12 interaction, and how the LEL (the Large Extracellular Loop) of Tspan12 interacts with Norrin, and that this loop is specific to Tspan12 (Tspan11’s LEL loop did not bind Norrin at concentrations tested). Adapted from Bruguera ES, Mahoney JP, Weis WI. eLife. 2025;13:RP96743. doi:10.7554/eLife.96743, used under CC BY 4.0.

[3-3] – Detergent choice for a GPCR is usually a stability question, and lauryl maltose neopentyl glycol (LMNG) is the most common choice by far, appearing in over half of all solved GPCR structures. Stability for a crystal structure is not the same thing as being in an active conformation. They purified a thermostabilized neurotensin receptor 1 variant into three detergents and measured agonist affinity on a Gator Prime. Instead of immobilizing the receptor, they loaded the small peptide agonist NT (8–13) onto Ni-NTA biosensors through a 6-His tag and titrated the receptor as analyte, which keeps the variable in solution. Affinity came in at 13.7 nM in decyl maltoside, 35.9 nM in dodecyl maltoside, and 94.9 nM in LMNG, meaning that ligand binding and receptor stabilization were, in this case, inversely proportional.

Figure 8, from Figure 2 in the above publication, showing how binding affinity changes depending upon membrane composition from least restrictive detergent structure (top) to most restrictive (bottom) for this GPCR. Adapted from Bower JB, van der Velden WJC, Gomez KP, et al, Protein Science. 2026;35(2):e70475. doi:10.1002/pro.70475, used under CC BY 4.0.

Running membrane proteins in nanodiscs or detergent? Our applications team can work through the immobilization strategy with you before you commit material.

Get in touch with us


[BACK TO TOP]

4 – Validating designed structures

Every study so far validated something that already existed in the lab. Structural prediction has advanced to the point that AlphaFold will model almost any complex you ask for, and then design tools like Boltz will hand you a binder for it. But until wet-lab data shows the interaction is real, it’s just a hypothesis.

AlphaFold has shown up several times earlier in this review: [1-1] paired AlphaFold3 with NLS mutagenesis, and [3-2] confirmed an AlphaFold Multimer interface through reciprocal charge swaps. Both aligned with the prediction. [1-3] shows the other outcome, where AlphaFold3 nailed the antibody fold to within 1 Å but placed the RNA nowhere near the real binding site.

De novo binder design is hit-or-miss by nature, which is why many AI-driven discovery programs generate thousands of binders in silico before they even touch a wet lab. The ones that survive often still number in the hundreds, and each one represents a potential data point for retraining and refining the AI algorithm. That’s why BLI, a high-throughput method that can measure kinetics and affinity without a purification step, is often the first wet-lab measurement a new binder sees. We mentioned a few AI-guided binder preprints in our previous blog post; the examples below add two more published works that used Gator BLI to validate de novo designs.

[4-1] – BindCraft is an open-source de novo binder development approach built on an iterative loop with AlphaFold2. The authors report that the method works without high-throughput wet-lab screening, citing experimental success rates from 10% to 100% (averaging 46%) across twelve targets. BLI was the first-pass method for validating these binder designs, with representative traces in Fig 2b, and it also carried the off-target analysis that checks whether a binder is merely sticky or selective. The team captured Fc-tagged PD-1, PD-L1, and IFNAR2, three immunoglobulin-like fold receptors that look alike structurally, on Gator Bio Protein A biosensors and dipped them into each binder at 1 µM. None of the binders cross-reacted.

[4-2] – Pathogenic E. coli steals iron by ripping the heme cofactor out of host hemoglobin. The transporter responsible, ChuA, grabs hemoglobin just long enough to strip the heme and let go, an interaction so transient that the team couldn’t capture the complex by crystallography or cryo-EM. BLI could measure it. With hemoglobin immobilized on Ni-NTA probes and ChuA titrated as analyte, they got a KD of 71.5 nM with fast on and off rates, which aligns with the bind-extract-release transporter hypothesis. Using AlphaFold, they identified extracellular loops 7 and 8 as possible binding sites for the interaction, and they used RFdiffusion and ProteinMPNN to design binders that would cover those loops. In total, they generated about 20,000 in silico and took their 96 top candidates to the lab for bacterial growth assays. Eight inhibited bacterial growth on hemoglobin agar. BLI put three of these inhibiting proteins at 127, 84.9, and 71.4 nM, while those same inhibitors showed no binding against three unrelated E. coli proteins, confirming the designs hit only their target.

Figure 9, from Figure 5f in the above publication, showing the KD of the growth-inhibiting proteins for ChuA. Adapted from Fox DR, Asadollahi K, Samuels I, et al, Nature Communications. 2025;16:6066. https://doi.org/10.1038/s41467-025-60612-9, used under CC BY 4.0.

If you are triaging designed binders and want the wet-lab step to keep up with the design step, tell us what you are expressing and at what scale.

Get in touch with us


[BACK TO TOP]

Final thoughts

These ten studies used Gator BLI to measure what the snapshot methods could not. ChuA grabs hemoglobin and lets go too fast for crystallography or cryo-EM to capture the complex. BLI measured it at 71.5 nM with fast on and off rates, which is what you would expect if the transporter binds, strips the heme, and releases.

No single Spike substitution was enough to let SARS-CoV-2 bind mouse ACE2. The virus needed both N501Y and E484K. To establish that, the team ran full titrations across five variants on a single plate.

Membrane proteins are harder. In [3-3] the team purified a neurotensin receptor into three detergents and measured a sevenfold range in agonist affinity, and the detergent that stabilised the receptor best gave the weakest binding. Measurements like that produce small shifts, and XT Optics gives you 3x signal to noise to read them.

Designed binders are the newest. In BindCraft and the ChuA work, BLI was the first wet-lab measurement anyone made on a new design. The antibody work in [2-2] is not de novo design, but it ran the same way. The team screened every substitution at one residue directly from clarified expression supernatant, with no purification step, on Strep-Tactin XT biosensors. Strep-Tactin XT is exclusive to Gator Bio.

These ten papers used seven capture chemistries on the same instrument, with no setup change between them. Those were Protein A, SA, Human Fc Gen II, Ni-NTA, Strep-Tactin XT, Amine Reactive and SMAP. That is what the biosensor menu gives you.

If you are working on a structural biology problem, our applications team can help you design the assay around it.

Get in touch with us


[BACK TO TOP]

References

[1-1] Bana J, Yariv A, Oppenheim T, et al. Importin 7 mediates the nuclear import of HIV-1 integrase via a specific interacting interface. FEBS Open Bio. 2026. doi:10.1002/2211-5463.70294

[1-2] Durie IA, Tehrani ZR, Karaaslan E, et al. Structural characterization of protective non-neutralizing antibodies targeting Crimean-Congo hemorrhagic fever virus. Nature Communications. 2022;13:7298. doi:10.1038/s41467-022-34923-0

[1-3] Bou-Nader C, Juma KM, Bothra A, et al. Structural basis of double-stranded RNA recognition by the J2 monoclonal antibody. Nature Communications. 2026;17:635. doi:10.1038/s41467-025-67414-z

[2-1] Ni D, Turelli P, Beckert B, et al. Cryo-EM structures and binding of mouse and human ACE2 to SARS-CoV-2 variants of concern indicate that mutations enabling immune escape could expand host range. PLOS Pathogens. 2023;19(4):e1011206. doi:10.1371/journal.ppat.1011206

[2-2] Goike J, Hsieh CL, Horton AP, et al. SARS-CoV-2 Omicron variants conformationally escape a rare quaternary antibody binding mode. Communications Biology. 2023;6:1250. doi:10.1038/s42003-023-05649-6

[3-1] Bruguera ES, Mahoney JP, Weis WI. Reconstitution of purified membrane protein dimers in lipid nanodiscs with defined stoichiometry and orientation using a split GFP tether. Journal of Biological Chemistry. 2022;298(4):101628. doi:10.1016/j.jbc.2022.101628

[3-2] Bruguera ES, Mahoney JP, Weis WI. The co-receptor Tetraspanin12 directly captures Norrin to promote ligand-specific β-catenin signaling. eLife. 2025;13:RP96743. doi:10.7554/eLife.96743

[3-3] Bower JB, van der Velden WJC, Gomez KP, Pan M, Bumbak F, Vaidehi N, Ziarek JJ. Stabilization versus flexibility: detergent-dependent trade-offs in neurotensin receptor 1 GPCR ensembles. Protein Science. 2026;35(2):e70475. doi:10.1002/pro.70475

[4-1] Pacesa M, Nickel L, Schellhaas C, et al. One-shot design of functional protein binders with BindCraft. Nature. 2025;646:483–492. doi:10.1038/s41586-025-09429-6

[4-2] Fox DR, Asadollahi K, Samuels I, et al. Inhibiting heme piracy by pathogenic Escherichia coli using de novo-designed proteins. Nature Communications. 2025;16:6066. https://doi.org/10.1038/s41467-025-60612-9

Blog

What is a Gator XT biosensor?

XT optics is what sets gator apart from the competition

Author: Benjamin Osborn, PhD; Anuja Patil, MS; Indrani Chakraborty, PhD


What is a BLI biosensor?

A biosensor is a detection instrument that uses a biological molecule to detect a specific target. These detector biomolecules (called “capture molecules” in this article) can include bacterial proteins like Protein A, antibodies/VHH and aptamers, each of which target molecules selectively. This allows biosensors to discriminate between samples that do contain the target molecule from samples that do not.

Biolayer Interferometry (BLI) is how selective binding is transduced into a meaningful signal. The inventor of BLI technology, Gator Bio’s CEO Hong Tan, discovered that when white light travels to the end of an optical fiber, a small layer of biomolecules (a biolayer) there reflects the light back into the fiber. A spectrometer can then observe that reflection.

A reflection isn’t a signal, but with the addition of a reference layer at the distal end of the biosensor, the two reflections (biolayer and reference) interfere with one another, changing the reflected light. The starting interference looks something like what’s pictured in Figure 1 in light grey. We already see some wavelengths amplified because the distance to and from the biolayer aligns with those wavelengths, creating constructive interference. Similarly, destructive interference occurs where that round-trip distance comes out to a half wavelength.

Figure 1: Exaggerated schematic representation of how antibody binding to a biosensor alters the interference spectrum in the visible range. Starting from a reference interference spectrum (grey), further biomolecule binding moves the peaks in the interference spectrum toward the infrared. This shift is plotted on a sensorgram as a function of time.

Figure 1: Exaggerated schematic representation of how antibody binding to a biosensor alters the interference spectrum in the visible range. Starting from a reference interference spectrum (grey), further biomolecule binding moves the peaks in the interference spectrum toward the infrared. This shift is plotted on a sensorgram as a function of time. Examples of sensorgrams are included in Figures 3 and 5.

As ligands bind to the surface, the peaks in the interference spectrum slide toward longer wavelengths (reds). Binding increases the optical thickness of the biolayer, which is its physical thickness multiplied by its refractive index change, so the reflected light completes a slightly longer round trip. The instrument reports that change in optical thickness in nanometers as the “nm shift.” Because optical thickness scales with the amount of material on the surface, the nm shift is a direct measure of how much biomolecule has accumulated.


How are Gator XT biosensors different?

XT biosensor optics is what sets Gator BLI apart from existing BLI technologies with dramatically improved sensitivity and performance.

When Dr. Tan revisited BLI technology and founded Gator Bio, he wanted to address a problem he had noticed early on in its development. He was getting far less signal than he had expected given the sizes of molecules that were binding to the surface. This is because traditional biosensor design captures less light from the biolayer than from the reference layer. While reduced sensitivity wasn’t a problem for a first-of-its-kind instrument that mostly focused on large molecule quantitation and binding kinetics, first-gen BLI biosensors clearly still had room for improvements.

So when Dr. Tan returned to the problem and founded Gator Bio, the Gator team revisited this discrepancy. Could our engineers capture more light from the biolayer? Could we improve performance of a BLI biosensor solely by improving the optics? Over the following years, the Gator team completely redesigned the surface of the biosensor, culminating in a proprietary process that improved the reflection from the biolayer back into the instrument. Without getting too technical, these improvements allow more of the light from the biolayer reflection into the optical fiber to reach the spectrometer, increasing the signal intensity (see Figure 2).

Figure 2: Schematic representation of a traditional biosensor (top) vs a Gator XT biosensor (bottom). Note how the total reflection from the protein layer (orange) has a much greater amplitude from XT biosensors which capture more of the reflected light, resulting in a much larger phase shift (nm shift). For this reason, XT biosensors give more signal per molecule

Figure 2: Schematic representation of a traditional biosensor (top) vs a Gator XT biosensor (bottom). Note how the total reflection from the protein layer (orange) has a much greater amplitude from XT biosensors which capture more of the reflected light, resulting in a much larger phase shift (nm shift). For this reason, XT biosensors give more signal per molecule

Now that the reflection from the biolayer is more intense, the resulting nm shift in the interference spectrum is approximately 3x greater relative to a traditional BLI biosensor. Figure 3 shows a head-to-head comparison of a traditional streptavidin BLI biosensor against a Gator SA XT biosensor featuring these new optical improvements. The difference in sensitivity and performance is clear just from this single experiment – a 3-4 nm loading shift was observed with a traditional BLI optical design, versus a 12 nm loading shift using Gator XT optics. This improvement was observable on the base model Gator Prime instrument (launched several years before any biosensors with XT optics were available), without requiring any redesign of the instrumentation or software.

Figure 3: Traditional BLI streptavidin sensors vs Gator Streptavidin XT. While the response is much greater on a Streptavidin XT biosensor, the antigen binding kon remains unchanged.

Figure 3: Traditional BLI streptavidin sensors vs Gator Streptavidin XT. While the response is much greater on a Streptavidin XT biosensor, the antigen binding kon remains unchanged.


Ligand Load (nm)
koff (s-1)
kon (M-1 s-1)
KD (M)
First-Gen SA
3.6
5.47*10-5
1.09*105
5.03*10-10
SA XT
12.5
8.47*10-5
1.01*105
8.42*10-10

Does increased sensitivity also mean increased noise?

Noise from biosensors with XT optics is still well beneath the instrument spec of 4 picometers of RMS noise, and the improvement to sensitivity dwarfs that small increase. As an example, across a panel of twenty-plus His-tagged proteins, we tested our anti-His XT (HIS XT) biosensor against both First-Gen BLI anti-his and our own traditional Gator Anti-His biosensor (See Figure 4 – full dataset here). The jump from First-Gen BLI to HIS XT is over 11x total signal on average. Moreover, some proteins did not bind to the First-Gen BLI’s biosensor at all, meaning that it’s not just a question of signal to noise, but whether you see anything at all.

Figure 4: Comparison in binding activity in a typical two-minute binding experiment across twenty different his-tagged proteins on each biosensor.

Figure 4: Comparison in binding activity in a typical two-minute binding experiment across twenty different his-tagged proteins on each biosensor.

Biosensor
Avg. signal loaded
vs. Legacy BLI
First-Gen Anti-His
0.17 nm
1x (some showed no binding at all)
Gator Anti-His
0.40 nm
2.4x
Gator HIS XT
1.93 nm
11.3x

That sensitivity comes from two upgrades in parallel: the XT optics themselves, plus a better capture molecule using Monod Bio’s anti-His NovoBody™ technology. Together, Gator’s HIS XT offers an 11.3x signal improvement without a meaningful increase in noise. Importantly: the signal gain from Gator XT biosensors comes by using improved optical physics and capture molecules, not sample rates and smoothing algorithms.

Why isn’t the noise increased as much as signal? A meaningful shift in the interference spectrum comes from a uniform layer of biomolecules across the surface, which produces a clean, consistent change in path length. A major source of noise, by contrast, is transient, non-uniform events at the surface and since these are incoherent, they couple weakly into the interference shift.

This is an important advantage of the XT format: increasing the gain from the biolayer amplifies the true binding signal without proportionally amplifying surface-level noise. The result is a stronger signal-to-noise ratio and better resolution of weak or low-level binding events. We’ll explore this mechanism in greater detail in a future article on the optical physics behind the XT platform.


Does XT optics allow you to study smaller analytes?

Although noise is incoherent, the signal amplification from XT optics does amplify coherent increases Figure 5 is a dataset with a 2 kDa peptide as analyte on a Gator Pivot instrument which used Peltier cooling to run the 16 channel assay at 20°C, 400 RPM with an SA XT biosensor. The XT sensitivity improvement results in a signal:noise ratio of over 200 – for a 2 kDa analyte!

Figure 5A: Full sensorgram showing binding by 150 kDa dimer as ligand and 2 kDa peptide as analyte. B: Analyte binding and kinetics fitting data. Analyte signal of 0.8 nm shift translates to over 200x signal:noise in a peptide binding experiment. XT optics increases response signal without also increasing noise.

Figure 5A: Full sensorgram showing binding by 150 kDa dimer as ligand and 2 kDa peptide as analyte. B: Analyte binding and kinetics fitting data. Analyte signal of 0.8 nm shift translates to over 200x signal:noise in a peptide binding experiment. XT optics increases response signal without also increasing noise.

In addition to the added sensitivity for smaller analytes, SA XT biosensors paired improved optics with a unique capture molecule design optimized for larger molecules. On SA XT, we use crosslinked streptavidin on the surface of the biosensor, removing the 3D matrix from the surface and that might hinder larger analytes, like lipid nanodiscs, from penetrating into the biosensor surface. This improved sensitivity and wider compatibility range across both larger and smaller molecules make SA XT a versatile troubleshooting sensor. It became one of our best sellers almost immediately. SA XT biosensors have been cited for numerous applications like nucleic acids [1], AI-driven protein engineering [2], and polysaccharide-protein interactions [3], among others. For more examples, you can search our citations page for Streptavidin XT or SA XT.


What are some other examples of XT biosensors?


Gator Bio’s XT optics innovation doesn’t increase the number of binding sites on the sensor – it generates more signal per molecule. This means that XT optics can be applied to any capture molecule on the biosensor. Gator has developed many different biosensors that take advantage of XT optics, so whatever your workflow, we likely have a biosensor with XT optics that meets your needs.

Strep-Tactin XT

Gator’s Strep-Tactin XT biosensor uses an engineered Streptavidin molecule, known as Strep-Tactin XT licensed from IBA Lifesciences, as a capture molecule to immobilize twin-strep-tagged proteins (tag sequence: SAWSHPQFEKGGGSGGGSGGSAWSHPQFEK). The twin-strep tag is two copies of the Strep-Tag II sequence (SAWSHPQFEK) which binds to the Strep-Tactin XT tetramer with picomolar affinity, making it an excellent choice for rapid capture and kinetics applications. They also regenerate up to 10 times, making them an excellent choice to run complex, multi-cycle assay formats. Note: proteins tagged with only a single copy of the Strep-tag II sequence will not remain tightly bound to Strep-Tactin XT. Gator offers an anti-Strep-Tag II antibody sensor for molecules that use a single Strep-Tag II tag.

Strep-Tactin XT biosensors originated from our custom biosensor program back in 2023. Built on the XT optical improvements, the Strep-Tactin XT biosensors can capture ligands from cell-free expression media, making them an ideal choice for AI-driven discovery workflows, where the XT optical improvements improve sensitivity for smaller targets. Recent AI-driven discovery and binder-characterization pre-prints have used Strep-Tactin XT in their workflows [4-6]. For more examples, you can search our citations page for Strep-Tactin XT.

Mouse Fc XT (MFC XT)

Mouse Fc XT also represents a significant improvement over traditional Mouse Fc biosensors. Not only does it have the same XT optical improvements, but it also includes a nanobody capture molecule with much faster on-rate and higher affinity, allowing faster screening of hybridoma pools for high throughput kinetics. Because MFC XT gives more signal per molecule, low-expressing clones that produce little or no signal on traditional Mouse Fc sensors often show a clear, positive signal on MFC XT, so you catch weak positives you would otherwise miss. For more information about the MFC XT biosensor, download this application note on using Mouse Fc XT biosensors for hybridoma screening.

Anti-His XT (HIS XT)

We’ve already talked quite a bit about HIS XT biosensors and their development journey in this previous blog post. HIS XT Biosensors combine XT optical improvements with Monod Bio’s new HIS XT NovoBody™, designed using AI to bind to a wide variety of His-tagged proteins with high affinity. HIS XT biosensors are also the best biosensor on the market for high throughput Fc receptor studies for two reasons:

  1. The capture molecule has no Fc (and therefore doesn’t cause any Fc interference).
  2. The sensitivity improvement derived from the XT optics (more signal per molecule) means that you can load fewer molecules on the surface, thereby improving the accuracy of kinetics experiments on this notoriously difficult interaction pair.

For a detailed protocol on how to use HIS XT for Fc receptor studies, download our application note studying CD32/FcγRIIa interaction here.

Regen SA (RSA)

Regen SA is our newest biosensor, employing an engineered streptavidin surface that can be completely regenerated using a simple two-step process. RSA delivers up to 4x higher signal and cuts cost per interaction by up to 8x versus non-reusable SA sensors. This makes it an ideal choice for high throughput screening for workflows featuring biotinylated molecules or instances where re-using an identical binding surface are important for consistency.

Other biosensors with XT optics

While many of our products are clearly labeled with the XT on the label, several of our other biosensors also incorporate XT optics into their biosensor design. Below is a complete list of biosensors that use Gator XT optics:

Biosensor
Capture Molecule
Key Advantage
Best Use Case
SA XT
Crosslinked streptavidin
Optimized surface for large molecules/complexes, sensitive enough for analytes as small as 2 kDa
Troubleshooting sensor; nucleic acids, AI-driven protein engineering, lipid nanoparticles (LNP, VLP, nanodisc)
Regen SA (RSA)
Engineered Streptavidin
Regenerates up to 10 times with our proprietary regeneration solution
Aptamers, Peptides, and biotinylated proteins of all types. Best with long-chain linkers (LC-LC; PEG4)
SMAP (Small Molecule and Peptide)
Streptavidin
High-density surface allows for visualization of small peptide and small molecule binders
Peptides, small molecules, and other low MW analytes
Strep-Tactin XT
Strep-Tactin XT (licensed from IBA Lifesciences), binds twin-Strep-tag
Regenerates up to 10x; compatible with cell-free expression media; lower ligand loading for cleaner kinetics
Cell-free capture, AI-driven discovery and binder characterization workflows
MFC XT
Anti-Mouse Fc nanobody (fast on-rate)
Faster screening of hybridoma pools; lower ligand loading for cleaner kinetics
High-throughput hybridoma screening; quality Mouse Fc kinetics
HIS XT
HIS XT NovoBody™, AI-designed by Monod Bio (no Fc domain)
No Fc interference; supports lower ligand loading for cleaner kinetics
Cell-Free capture; Fc receptor studies (e.g., CD32/FcγRIIa); His-tagged protein kinetics
Anti-VHH
Polyclonal anti-VHH cocktail
Binds to no-tag VHH; supports capture from cell-free media
High throughput screening
Anti-human FAB
Anti-FAB antibody
Binds to untagged human FAB fragments
FAB screening, FcRn
AAVX/AAV9
CaptureSelect AAVX/AAV9 nanobody
Binds only to intact capsids, giving viral particle titer
Upstream/downstream viral titer determination
Anti-PEG
Anti-PEG antibody
Binds to PEG of any length, capture LNPs with PEGylated lipids
LNP characterization for selective CAR-T mRNA delivery
Amine Reactive (AR)
Classical label-free chemistry
Increased sensitivity using amine-reactive format
No-tag label-free experiments, reduced NSB for C1q assays over First-Gen BLI

In addition to the list above, custom biosensors can be made with XT optics on request. For any questions about our custom biosensors program contact our team.


Should I approach my experiments differently with XT optics?

The same best practices of BLI still apply. However, XT biosensors give you a leg up in four different ways:

  1. Common best practice is to load as few ligand molecules as possible while obtaining a usable amount of signal. With Gator XT optics amplifying signals by 3x, you can load to the same nm shift as you are accustomed to and thereby load 3x fewer ligand molecules onto the surface without affecting analyte binding sensitivity. This can have a dramatic impact for applications that are affected by mass transport or avidity, like Fc receptors (see the His XT application note on Fc receptor studies).
  2. Increased sensitivity also gives you flexibility to study analytes down to 2 kDa. Load as much ligand as is needed to observe peptide binding clearly.
  3. For picomolar binders, a greater measurable amount of dissociation is helpful for calculating koff. With an XT biosensor, you can observe slow off rates more clearly because more signal is being removed from the sensor surface through dissociation, even if it’s the same number of molecules. This can improve accuracy of the modeling software as it tries to fit slow-off kinetics.

How can I compare the difference between XT and standard BLI biosensors in my own lab?

Gator’s XT biosensors are only compatible with Gator instruments, and not with any competitor’s instrumentation. However, if you’d like to see how Gator XT biosensors could improve your data quality, contact us and our team can arrange a demonstration in your lab.


Works Cited

[1] Lee, C.-C., Hsu, S.-F., Chang, Y.-W., Ho, M.-R., Sugiyama, H., Wang, W.-C., & Wang, A. H.-J. (2026). Structural basis of B-to-Z DNA transition mediated by an anti-Z-DNA antibody. Nucleic Acids Research, 54(5). DOI: 10.1093/nar/gkag160. Full text: academic.oup.com/nar/article/54/5/gkag160

[2] Zhu, Y., Liu, H., Qu, F., Qu, C., Wang, Y., Yao, J., Li, K., Wang, Z., You, S., Hua, L., Ge, C., Yao, H., & Li, T. (2025). De novo binders overcome the MMLV RT stability-activity trade-off. iScience, 28(12), 114140. DOI: 10.1016/j.isci.2025.114140

[3] Xiao, R., Zhou, W., Peng, J., Cai, S., Xie, J., Liao, W., Li, S., Ding, K., & He, F. (2025). AJDW, polysaccharide from Albizia julibrissin, binds to PI3K and induces cell cycle arrest and ROS-mediated pancreatic cancer cell apoptosis. Glycoscience & Therapy, 100003. DOI: 10.1016/j.glycos.2025.100003

[4] Stark, H., Faltings, F., Choi, M., Xie, Y., Hur, E., O’Donnell, T., Bushuiev, A., Uçar, T., Passaro, S., Mao, W., Reveiz, M., Bushuiev, R., Pluskal, T., Sivic, J., Kreis, K., Vahdat, A., et al. (2025). BoltzGen: Toward Universal Binder Design. bioRxiv. DOI: 10.1101/2025.11.20.689494

[5] Du, Y., Yu, B., Liu, T., Shen, T., Chen, J., Rittig, J. G., Sun, K., Zhang, Y., Song, Z., Zhou, B., Masschelein, C., Wang, Y., Wang, H., Jia, H., Zhang, C., Zhao, H., Ester, M., Head-Gordon, T., Gomes, C. P., Sun, H., Duan, C., & Schwaller, P., Jin, W. (2025). Accelerating Scientific Discovery with Autonomous Goal-evolving Agents. arXiv:2512.21782. Link: arxiv.org/abs/2512.21782

[6] Candido, S., Hayes, T., Derry, A., Rao, R., Lin, Z., Verkuil, R., Wu, B. Z., Lee, J. S., Bruguera, E. S., Keval, J. A., Kopylov, M., Pak, J. E., Wu, W., Thomas, N., Mataraso, S., Rives, A., et al. (2026). Language Modeling Materializes a World Model of Protein Biology. bioRxiv. DOI: 10.64898/2026.06.03.729735

Blog

HIS XT Biosensor: How AI-Designed Protein Engineering Solved the Anti-His Baseline Drift Problem

If you’ve ever struggled with baseline drift, slow loading times, or unreliable kinetics when working with His-tagged proteins, you’re not alone. These are longstanding pain points that have affected every anti-His biosensor on the market. In this post, we pull back the curtain on how Gator® Bio partnered with computational protein design company Monod Bio to develop HIS XT, a next-generation anti-His biosensor built around an AI-designed NovoBody™ protein. From the fundamental limitations of monoclonal antibodies to the in silico design process that replaced them, we walk through exactly why traditional anti-His probes fall short for BLI and how we engineered our way around those limitations to deliver faster loading, dramatically reduced baseline drift, and consistent performance across N- and C-terminal His-tagged proteins.

Behind the Scenes

Why Did Researchers Need a Better Anti-His Biosensor?

The humble His tag (in its many variants) is likely the most common protein tag in existence, but as any user knows, its simplicity can sometimes come at a cost. In this post, we’ll walk through our multi-year journey to develop a better anti-His biosensor, starting from the fundamental problems we encountered to how we solved them by collaborating with AI protein design experts to generate the best possible anti-His biosensor on the market. Whether you use Gator instruments or not, we hope the technical insights are useful for anyone working with His-tagged proteins.

Our traditional anti-His biosensor is used by hundreds of customers around the world. His-tagged proteins are ubiquitous in research for good reasons: expression plasmids often include the tag by default, the tag is small and typically doesn’t affect protein structure or activity, and purification resins from multiple manufacturers help keep costs competitive.

Anti-His was one of our first biosensors because it allowed users to capture His-tagged proteins directly from crude lysate and obtain quantitation or kinetic information without complete purification. For many labs, this workflow represents a significant time savings. Even after we launched Ni-NTA biosensors, which have slightly better affinity for His-tagged ligands, anti-His remained the more popular choice.

However, customer feedback revealed consistent pain points:

  • Loading steps often required 5 minutes or longer
  • Baseline drift interfered with accurate kinetic measurements
  • Weak ligand binding signal made picomolar interactions with smaller analytes difficult
  • Performance varied significantly between different His-tagged proteins

These weren’t Gator-specific problems. Every BLI vendor’s anti-His probe suffered from similar issues because they all relied on the same fundamental approach: monoclonal antibodies. While Gator’s anti-His biosensor outperformed other commercially available options in baseline stability, we knew we could do better.


Sidebar: About Anti-His and Baseline Drift

What is Baseline Drift, and Why is Anti-His So Problematic?

Anti-His biosensors use a monoclonal antibody to hold His-tagged ligands on the biosensor, but those antibodies have notoriously weak interactions. When the capture molecule slowly dissociates from the biosensor surface during measurement, we call the resulting downward slope “baseline drift” because we still see that downward movement on the baseline sensorgram. This interferes with accurate kinetic analysis because although we can subtract that baseline from the analyte binding data, the subtraction isn’t perfect – the analyte binding step contains analyte associating and dissociating from the ligand, as well as ligand dissociating (both with and without analyte attached). In many cases, this subtraction introduces large fitting errors that make accurate analysis difficult.

This is particularly problematic for tight-binding protein interactions with slow dissociation rates. When the capture antibody-his-tagged ligand interaction has a faster off-rate than the his-tagged ligand-analyte interaction being measured, it becomes impossible to distinguish his-tagged ligand dissociation from analyte dissociation, compromising data quality.

two sensorgrams side by side comparing Gator Bio's new HIS XT biosensor with other commercially available anti-His biosensors

Figure 1: Gator anti-His vs another commercially available anti-His biosensor. Both sensors were loaded with his-tagged PD-L2 to ~0.5 nm, then both sensors were allowed to dissociate for 1800 seconds. In that 1800 second baseline, Gator’s traditional anti-His lost 18% of its binding signal, while the alternative lost 62% of bound protein.

Why Can’t Traditional Anti-His Probes Measure Picomolar Interactions Reliably?

Anti-His antibodies typically have relatively weak affinity for His-tags (micromolar to nanomolar range), which means they have fast off-rates (koff). When measuring picomolar protein-protein interactions that dissociate very slowly, even small amounts of probe dissociation create baseline artifacts that obscure the true signal. By using tight-binding capture molecules with slower off-rates, you’re more likely to accurate picomolar kinetics measurements.


Making HIS XT a Truly Next-Generation Design

What we looked for as we used computational protein design to solve persistent challenges in His-tag capture.

What Makes a Good BLI Capture Molecule?

Antibodies have been the standard capture molecules for decades because they’re straightforward to generate, commercially available, and work adequately for many applications. However, when we examined what makes an ideal BLI capture molecule specifically, we identified several inherent limitations with antibody-based approaches. They are:

  1. Weak binding affinity
  2. Inconsistent performance across proteins
  3. Avidity
  4. Regeneration efficacy

Over several years, we evaluated dozens of anti-His antibodies from different manufacturers, hoping to find one with better baseline stability and faster association rates. What we discovered was that antibodies optimized for traditional immunoassays often lack the characteristics needed for high-performance BLI.

Why are Anti-His Monoclonal Antibodies Consistently Weak Binders?

There are three general reasons why anti-His antibodies aren’t the strongest:

  1. Manufacturers of monoclonal anti-His antibodies don’t have the same requirements that a BLI biosensor manufacturer does. Many anti-His manufacturers optimize for elution or stability, making the reagent good enough for a western and not a label-free technique. Gator biosensors, on the other hand, need a slow dissociation rate.
  2. Anti-His antibodies target a repeating motif; this means that there’s a high likelihood that an interaction can be “off-center” and weakly-binding when first presented with the tag. Additionally, many antibodies have a strong preference for C-terminal His tags because that helps with achieving a proper binding configuration, but this comes at the cost of versatility.
  3. Antibodies against His tag are hard to make because the tag is poorly immunogenic: it’s short, it can have variable charge depending on its environment, and it can be difficult to access when conjugated to a protein.
Why Do Anti-His Antibodies Have Variable Affinity to Different Proteins and Tag Locations?

Anyone who has shopped for antibodies knows that manufacturers rarely advertise kinetic parameters—the kon, koff, or KD of the antibody-antigen interaction. For most antibodies targeting consistent protein epitopes, this information would be relatively straightforward. However, with anti-His antibodies, achieving consistent binding across different proteins is extremely difficult.

We generated single-point KD and koff estimates for our traditional anti-His and the new HIS XT biosensors across 20 different His-tagged proteins from 5 manufacturers. The variation was substantial. This is because the tags are in different electrochemical environments, some have likers (like G4S) between tag and protein, and the tags can be at either the C or N-terminus. When the tag has such low variability, the surroundings can play a large role in availability.

Figure 2 contains the KD and koff of each of these tested proteins. We observed a relatively wide variability in both KD and koff rate with traditional Gator anti-His biosensors, which is dramatically improved with HIS XT. To end users, this means that HIS XT will be far more reliable, across a variety of experimental candidates.

Two graphs depicting koff and KD estimates of 20 different his-tagged proteins for Gator Bio traditional Anti-His and new HIS XT biosensors

Figure 2: koff and KD estimates of 20 different his-tagged proteins to both the Gator anti-His and HIS XT biosensors. Results are single-point experiments with results generated via curve fitting. Anti-His has only 19 results plotted above because one failed to bind altogether. Note that the HIS XT biosensor shows a consistent improvement in overall binding, as well as interaction longevity, with far fewer instances of poor binding.

What is Avidity? Why Would a Monoclonal Antibody as a Capture Ligand Cause This Problem?

In the context of BLI, avidity usually refers to having two proteins on the surface of a biosensor interacting with a single protein in solution. When you use a bivalent capture molecule on the surface of the sensor, you have two capture sites in immediate proximity, meaning that captured ligands could be close enough to each other to “double-bind” to an analyte, or to prevent analyte from escaping because another binding site is immediately present. 

These artifacts are extremely common on BLI, and make the ligand-analyte interaction appear much tighter than actual 1:1 affinity, complicating accurate kinetic measurements. By using a small single-domain binding protein as capture molecule, we limit this issue.

What is Regeneration? Why is Regeneration Performance Important for Biosensor Development?

Regeneration efficacy (the ability to withstand short exposure to low pH conditions and successfully remove bound His-tagged protein) is critical for both practical and economic reasons. Being able to reuse the same biosensor for multiple binding cycles not only reduces cost per assay but also ensures consistency. Poor regeneration means either replacing biosensors mid-experiment or accepting incomplete ligand removal (which appears as inconsistent load and inconsistent analyte binding because bound ligand isn’t completely removed). For high-throughput applications, being able to run the entire plate (typically 12 columns) on a single sensor can dramatically simplify the workflow.

Even after narrowing our list to a handful of high-performance, high consistency candidates, regeneration efficacy was one of the most important discriminators between the final candidates. Similar to the kinetics experiments above, we wanted to ensure that the final candidate for the new HIS XT would consistently work well for a variety of possible ligands. Figure 3 shows some examples of this process from the final candidate.

Four sensorgrams depciting regeneration efficacy of Gator Bio's new HIS XT biosensor across 4 different his-tagged proteins

Figure 3: Regeneration efficacy of His XT across four different his-tagged proteins. Top-left: His-HSA; top-right, PD1-his as ligand, Anti-PD1 as analyte; Bottom-left: MD1-his as ligand, PD-L2 as analyte; Bottom-right: Human Thrombopoietin Receptor-His as ligand, Human Thrombopoietin as analyte. Note: Regeneration efficacy may vary by protein and the chemistry surrounding the tag.


An in silico Approach

After years of screening antibodies from various suppliers, we reached a conclusion: there simply wasn’t a better antibody available. Gator’s traditional anti-His biosensor was already best-in-class among commercially available options, but it still had the fundamental limitations inherent to antibody-based capture.

The breakthrough came from a different approach entirely. We partnered with Monod Bio, a company specializing in computational protein design. Their AI-driven methods had generated proof-of-concept results suggesting they could create a superior anti-His binding protein.

How Did AI Design Accelerate Biosensor Development?

Rather than relying on biological evolution (as in antibody generation from immunized animals) or combinatorial libraries (as in phage display), the only way to generate a truly new solution for anti-His binding was through an in silico approach. Using AI for protein design starts with the desired binding characteristics and works backward to design a protein sequence that should achieve them.

For His-tag capture, we asked Monod Bio to explicitly optimize for strong binding to the poly-His motif at N- and C-terminal tags, single-domain architecture to avoid avidity, and stability under regeneration conditions. This approach bypassed months of screening and provided greater consistency for both association and dissociation kinetics.

What are Monod Bio’s NovoBody™ Proteins?

Rather than stick with the traditional antibody and nanobody scaffolds already on the market, Monod Bio moved in a new direction, creating NovoBody™ proteins, de novo computationally designed binding proteins. Unlike antibodies derived from animal immunization or VHH domains from camelids, NovoBody™ proteins are designed from scratch using AI algorithms. They can be optimized for specific requirements. In this case, they were designed for consistent His-tag binding, single-domain structure, and regeneration stability. Since they’re single-domain NovoBody™ proteins avoid the avidity complications inherent to bivalent antibodies.

Over the course of a year, Monod Bio sent us several candidates as they refined their designs and we optimized our conjugation chemistry. Many candidates showed acceptable kinetics, but some had difficulty conjugating to biosensor surfaces, others showed inconsistent regeneration performance, and ultimately only one emerged as clearly superior.

As part of our validation process, we tested 27 different His-tagged proteins representing a range of molecular weights, tag locations (N- and C-terminal), charge characteristics, and manufacturers. This panel allowed us to assess the pan-specificity of HIS XT, and the data confirmed dramatically improved binding activity, stability, and consistency compared to traditional antibody-based probes. Figure 4, below, shows the result, as well as the comparison against both anti-His and other ommercially-available anti-his biosensors on the market.

Two sets of graphs comparing load shift and loss in a sample experiment across 20 different his-tagged proteins

Figure 4: Loading shift and loss in a sample experiment across 20 different his-tagged proteins of varying sizes, tag locations, and manufacturers.

Why Call It HIS XT? What Does XT Mean for Gator Bio Biosensors?

While the capture molecule is critical, biosensor performance also depends on surface chemistry and optical design. Biosensor development also requires immobilizing that capture molecule, optimizing its surface chemistry and tailoring the optical properties of the biosensor.

One key innovation is preparing the biosensor using our proprietary XT optical coating, which dramatically increases the total signal per molecule on the surface of the sensor. The number of molecules on the surface at 1 nm shift on HIS XT is much lower than the number of molecules on a traditional anti-His at the same signal shift. This means you can still obtain usable data with smaller analytes without having to overload the biosensor for sensitivity. It also means that you can load far fewer proteins onto the surface and maintain good signal, reducing the risk of avidity further.

As previously shown in Figure 4, in standardized 2-minute experiments, HIS XT sensors generated over 4× the total signal compared to other commercially available anti-His antibody biosensors when using the same ligand loading well. This is due to both an improved kon association rate, as well as this amplification of total signal.

Technical Considerations for Optimal Performance

As with any high-performance capture system, there are protocol optimization considerations to keep in mind. To experienced users, these points are nothing new, but it’s important to bear them in mind when optimizing your protocols.

  1. Beware of over-loading your biosensors
    Because HIS XT biosensors bind His-tagged proteins so efficiently, it’s easy to overload them using conventional protocols. For kinetic assays, we recommend using 10 nM of protein or less. The higher signal per molecule means less protein is needed to achieve excellent signal-to-noise ratios.
  2. Managing second-baseline dissociation
    Although the anti-His NovoBody™ remains bound with a very long interaction half-life, some ligands will dissociate a small amount at the beginning of second baseline (Figure 5). Using less protein during loading and avoiding saturation helps limit this effect. It is a good idea to include a 60-120 second baseline after loading allows this initial dissociation to plateau, though with proper reference subtraction this may not be necessary.
  3. Optimize regeneration conditions
    We screened the capture molecule used on the HIS XT biosensor for regeneration efficacy, and can confidently say that when using our Regen Buffer pH 2.0 (No Salt) (PN: 120008), you can typically expect 10+ regenerations with the HIS XT biosensor. However, His-tag chemistry can vary, and some proteins may require optimization of regeneration conditions beyond low-pH buffer wash.

    We do not recommend regenerating with low-pH buffers that contain high salt, as this may reduce the effectiveness of the biosensor.
Sensorram depicting the anti-His NovoBody™ maintaining a long binding.

Figure 5: HIS XT biosensor loaded with his-tagged PD-L2, then allowed to dissociate for 1800 seconds. In that 1800 second baseline, only 9% of bound protein loss was observed, most of which occurred within the first 60-120 seconds.


Frequently Asked Questions


Conclusion

After years of optimization and development, we’re proud to introduce the HIS XT biosensor. We believe it represents a significant advancement in His-tag capture technology, addressing the many fundamental limitations that have persisted across the field for years.

Have you experienced similar challenges with His-tagged proteins and capture in your own work? We’d love to hear about your applications and learn what other capture molecule challenges the research community is facing. Share your thoughts via email to [email protected].


Getting Started with BLI

Ready to implement BLI in your laboratory? Here’s how to get started:

  1. Define your application: Determine whether you need quantitation, kinetics, or both
  2. Select your biosensor: Choose the immobilization strategy that fits your molecules
  3. Optimize your assay: Work with our applications team to dial in conditions
  4. Scale up: Implement automation for high-throughput workflows

Request Quote or Demo

Blog

BLI Basics – What is BLI and How Does it Work?

Understanding how proteins interact is fundamental to biologics development. Biolayer interferometry (BLI) makes these invisible interactions visible with real-time, precise visualization of binding interactions.

What is BLI?

Biolayer interferometry (BLI) is an optical, label-free technique that measures real-time binding at a sensor surface. BLI captures binding events as they happen, so it can measure both the rates and extent of molecular interactions, giving it broad appeal for quantitative kinetic and affinity characterization throughout biologics discovery and development.

BLI works by analyzing interference patterns created when white light passes through specialized biosensors. Here’s how the technology works:

The instrument shines white light through a fiber optic cable and into a glass biosensor. The biosensor tip is functionalized with capture molecules that bind target proteins of interest when placed into solution. As proteins bind and the layer of bound proteins grows, the instrument detects that change in distance through optical interference between two reflection points: a reference layer near the sensor surface and the outer edge of the protein layer.

A spectrometer continuously monitors these reflections. After establishing a baseline measurement, the system tracks how the interference pattern shifts across the entire visible spectrum as the protein layer changes thickness. The instrument plots these changes on a graph called a sensorgram (a real-time binding curve), with time on the X-axis and “nm shift,” the measurable change in interference pattern, on the Y-axis.

Watch how BLI translates molecular binding into actionable data


Gator Bio’s BLI Platform

Gator Bio’s BLI instruments allow users to analyze samples in standard 96-well or 384-well microplate formats, measuring up to 32 samples simultaneously. The plate-based design enables the instrument to move biosensors sequentially from well to well, capturing binding data at each step of your assay protocol. This high-throughput capability means you can screen hundreds of samples per day while maintaining the kinetic precision needed for confident decision-making.

Explore Gator Bio BLI Instruments


How BLI Measures Binding

BLI transforms binding responses into quantitative data through two primary approaches: measuring binding kinetics or measuring total binding response.

Quantitation Workflows

For concentration measurements, BLI monitors the initial rate of binding (nm shift per second), which is directly proportional to both the association rate constant (kon) and analyte concentration. This relationship creates a concentration-dependent signal: higher concentrations produce steeper initial slopes.

To quantify unknown samples, you run a series of known standard concentrations to generate a calibration curve, then compare your unknowns against this standard curve. Because the analysis focuses on initial binding rates, these quantitation experiments are remarkably fast, often finished in less than 2 minutes per sample.

Rapid quantitation through initial rate analysis delivers results in under 2 minutes per sample

Kinetics Workflows
For kinetic characterization, BLI provides real-time measurement of both binding (association) and unbinding (dissociation) between biomolecule pairs. This allows you to determine three critical parameters through curve fitting: – kon (association rate constant) – how fast molecules bind – koff (dissociation rate constant) – how fast molecules unbind – KD (equilibrium dissociation constant) – overall binding affinity.

For interactions that reach equilibrium quickly, you can use endpoint binding measurements to determine KD . The minimal setup time and multi-channel format make BLI particularly attractive for high-throughput screening campaigns where you need to evaluate dozens or hundreds of candidates.

Real-time kinetics reveal the complete binding profile—from initial association through full dissociation



Key Terms in BLI

Understanding BLI terminology helps you read and understand BLI literature:

  • Capture Molecule: The reagent coated on the probe surface that is specifically designed to capture the first molecule of the binding pair
  • Ligand: The first molecule of the binding pair captured onto the biosensor surface
  • Analyte: The second molecule that binds to the immobilized ligand (your molecule of interest)
  • Baseline: A buffer-only step that establishes the sensor’s stable starting point, allowing you to distinguish meaningful binding signal from instrument drift. A second baseline can also help to verify ligand binding to the biosensor.
  • Loading: The kinetics assay step where biosensors are exposed to ligand-containing solution, allowing ligand to immobilize on the sensor surface
  • Association: The kinetics assay step where analyte-containing solution contacts the ligand-loaded biosensor, enabling binding to occur
  • Dissociation: The kinetics assay step where the biosensor returns to buffer-only conditions, allowing bound analyte to release from the sensor

Assay Design Considerations

Successful BLI experiments require a thoughtful experimental design. Here are the most critical considerations:

Buffer Matching is Critical

BLI is an optical technique that responds to changes in the refractive index. Variations in buffer composition between assay steps can create signal artifacts that overwhelm your binding data. For optimal performance:

  • Maintain consistent buffer conditions across all assay steps
  • Watch for refractive index shifts from pH changes, DMSO, glycerol, sucrose, and other solvents
  • Remember that glycerol isn’t volatile and may remain in lyophilized proteins
  • For quantitation assays, pre-soak biosensors in buffer similar to your samples to minimize non-specific signal.
Choosing the Right Immobilization Strategy

Kinetics assays require immobilizing your ligand to the biosensor without blocking the analyte binding site. The most common strategies target conserved domains or protein tags:

Common capture approaches:

  • Fc-binding biosensors: Protein A, Anti-Mouse Fc XT, Anti-Human Fc for antibodies
  • Tag-based biosensors: Anti-His, Ni-NTA for His-tagged proteins; Strep-Tactin XT for twin-Strep-tagged proteins
  • Biotin-Streptavidin: non-reversible, Streptavidin biosensors for biotinylated proteins (note: use a biotinylation protocol that limits the number of biotins per molecule – lysines are common in binding domains, and biotinylation can dramatically reduce analyte binding efficacy)

Fc-binding and Tag-based biosensors offer the advantage of regeneration: a simple low pH wash typically removes bound analyte, allowing biosensor reuse and reduced costs.

Browse Gator Bio Biosensors

Optimizing Analyte Concentration Range

Proper concentration selection is essential for accurate kinetic measurements. Your concentration series should:

  • Upper range: Highest concentrations should show clear curve saturation (beginning to plateau)
  • Lower range: Lowest concentrations should remain well below half-maximal binding
  • Spacing: Use serial dilutions (typically 2-fold or 3-fold) to cover 2-3 orders of magnitude
Concentration too high
Concentration too low
Optimal concentration

Simulation of 1:1 binding interaction using Gator software’s built-in simulation feature for a 1 nM KD interaction, using 1 nm of response to represent 100% saturation of the available binding space in the simulation. Presented are several concentration series, each with a two-fold serial dilution of analyte. The sample labeled “too high”  begins at 1000 nM, the “too low” sample begins at 2 nM, and the “optimal” sample begins at 50 nM.

Minimize Ligand Loading

A common mistake is overloading the ligand on the biosensor. More isn’t better—excessive ligand density can cause:

  • Avidity effects that complicate analysis
  • Reduced biosensor regeneration efficiency

Use the minimum ligand loading level that still provides clear analyte binding signal. This typically means aiming for 0.5-1.0 nm loading signal for most antibody-based assays.


Where BLI Fits in Your Analytical Toolkit

Transitioning from ELISA

Many laboratories upgrade from ELISA to BLI because real-time binding visualization provides complete kinetic context that endpoint assays simply can’t match. Instead of a single data point, you see the entire binding and unbinding process unfold. This makes troubleshooting straightforward; if something isn’t working, the sensorgram tells you exactly where the problem lies (poor binding, rapid dissociation, non-specific interactions, buffer effects).

Key advantages over ELISA: – Real-time kinetic information, not just endpoint binding – No washing steps required – Label-free detection – Direct observation of binding specificity – 10-100x faster assay development

Complementing SPR Workflows

SPR laboratories often employ BLI for rapid candidate screening and prioritization. While SPR excels at detailed characterization of a few high-priority candidates, BLI’s multi-channel format enables you to quickly evaluate hundreds of candidates and identify the top 10-20 for deeper analysis.

Why labs add BLI alongside SPR:

  • Screen hundreds of samples per day (even base model instruments)
  • Wider variety of off-the-shelf biosensors for diverse project types
  • Plate-based format simplifies automation integration
  • Faster setup reduces time from sample to data
  • Lower cost per sample for routine screening
Quality Control and Bioprocessing Applications

BLI has become indispensable in manufacturing settings, delivering:

  • Rapid quantitation: Real-time titer measurements that drive process decisions
  • Potency assays: Activity measurements that meet regulatory requirements
  • Clone selection: High-throughput screening during cell line development
  • Formulation stability: Rapid assessment of protein aggregation and degradation

In quality control environments, BLI’s speed and reproducibility make it ideal for routine lot release testing and in-process monitoring.

Explore BLI in Manufacturing


Quick Comparison: Gator BLI vs Other Binding Technologies

FeatureGator Bio BLIELISASPR
Real-Time Kinetics✓ Yes✗ No✓ Yes
Label-Free✓ Yes✗ No✓ Yes
ThroughputHigh (32 Channels)Medium-HighLow (1-8 channels)
Setup Time< 15 mins2-4 Hours30-60 min
Automation-Ready✓ Plate-Based✓ Plate-BasedLimited
Biosensor VarietyVery HighN/AMedium
Sample ConsumptionLow (μL)Medium (μL-mL)Very low (μL)
Best ForScreening & QCEndpoint AssaysDetailed characterization


BLI: The Workhorse of Bioanalytical Labs

BLI has earned its place as the workhorse platform in bioanalytical laboratories because it delivers the binding and quantitation data you need with unmatched speed and versatility. Whether you’re discovering new therapeutics, optimizing manufacturing processes, or ensuring product quality, BLI provides the analytical foundation for confident decision-making. 


Getting Started with BLI

Ready to implement BLI in your laboratory? Here’s how to get started:

  1. Define your application: Determine whether you need quantitation, kinetics, or both
  2. Select your biosensor: Choose the immobilization strategy that fits your molecules
  3. Optimize your assay: Work with our applications team to dial in conditions
  4. Scale up: Implement automation for high-throughput workflows

Request Quote or Demo

Blog

What is Biomolecular Interaction Analysis?

Biomolecular interaction analysis seeks to understand the interactions between biomolecules and the results of these interactions. This type of analysis is critical for medical research and pharmaceuticals. Here we explain some of the key principles of biomolecular interaction analysis and its applications.

Why Study Biomolecular Interaction?

Interactions between biomolecules are a critical aspect of many biological processes. Whilst some proteins function on their own, most proteins are only active in complex forms. This means that their relationships with other biomolecules such as metals, lipids, proteins, and nucleic acids are extremely important.

Biomolecular interaction analysis looks at the physical contact between protein partners as well as the functional trends in signaling and metabolic pathways. Protein interactions and molecular interactions are intrinsic to all areas of cellular process such as:

  • Regulation of metabolic pathways
  • Cellular motion
  • Signal transduction
  • Environment sensing

With so many critical processes being informed by biomolecular interaction, the right form of analysis must be chosen.

Surface Plasmon Resonance SPR

SPR is a label-free method of biomolecular interaction analysis that works by using incident light to stimulate oscillating conduction electrons between positive and negative permittivity material. SPR measures adsorption onto planar metal or onto the surface of metal nanoparticles.

SPR is well-suited to measuring the affinity and selectivity of biomolecular interactions. It can be used for investigating association and dissociation rate constants and presenting the kinetics of biomolecular interactions. This technique can also be used for equilibrium binding analysis.

SPR  does not require additional reagents, sample preparation, or assays and has high precision for classifying protein-protein interactions.[1]

Isothermal Titration Calorimetry

ITC is a method used for the quantitative study of a range of biomolecular interactions. This technique works by measuring the heat directly that is absorbed or released in a biomolecular binding event.

The main benefit of ITC is that it can determine each binding parameter concurrently in just one experiment and no modification of binding partners is required.

Bio Layer Interferometry BLI 

BLI (biolayer interferometry) is an optical biosensing technology used for biomolecular interaction analysis. One of the key benefits of BLI is that it offers real-time label-free measurements for data analysis.

BLI studies macromolecular interactions, focusing on the interference patterns from white light reflected from the surface of the biosensor tip. It is a simple technique that measures Kon, Koff, and KD in just one rapid assay.

BLI is robust and easy to set up, it can be used for a range of samples including both crude and purified samples. This means that it can be used for a variety of investigations without needing to swap between systems for samples. Its small footprint means that it does not take up valuable space in the lab, making it more versatile than many other technologies.

BLI also allows for a broad variety of affinity values with the ability to investigate the following:

  • Nanobodies
  • Antibodies
  • AAV
  • Small molecules
  • Nucleic acid
  • Membrane proteins

Biomolecular Interaction Analysis from Gator Bio

Gator Bio stocks a range of BLI models suitable for biomolecular interaction analysis. It is a highly flexible, accessible, and sensitive technology that can characterize large varieties of proteins and molecules. If you would like to find out more about how our products could help with your biomolecular interaction analysis, read our dedicated page here.

[1] https://pubmed.ncbi.nlm.nih.gov/23533209/

Blog

Why are Binding Kinetics Important?

Binding kinetics have been studied for many years and make up the basis of pharmacological theory today. Whilst all pharmaceutical drugs are different, the binding kinetics have many similarities. 

The term binding kinetics refers to how quickly a compound binds to its target and the speed at which it dissociates from it. Binding kinetics measure the on-rate and the off-rate but have historically been challenging to measure. This article will explain the fundamentals of binding kinetics and how they work. 

The Importance of Binding Affinity 

The binding affinity of a drug is inextricably linked to binding kinetics. The binding affinity refers to the levels of drugs needed to occupy 50% of the target molecules at equilibrium. This number is the target occupancy, used to predict in-vivo efficacy.  

The residence time is a quantitative value referring to how long it takes for drug-receptors to reach target occupancy.  

The Importance of Binding Kinetics in Affinity Measurements 

The importance of binding kinetics is based on the fact that the time to reach equilibrium is dependent on the dissociation time. When the opposing rates are equal, a binding interaction is at equilibrium. Because association takes place first for dissociation to occur. This is why full binding kinetics data is required to effectively measure binding affinity. When using binding kinetics, meeting an equilibrium isn’t required and therefore it can be a much faster method. 

Binding Kinetics in Drug Discovery 

Accurate and fast measurement of binding affinity is critically important in the drug discovery process. If the actual affinity is above what is measured then patients may risk overdose in clinical drug trials. Similarly, if it is lower than recorded potentially life-saving drugs could be missed because of under-dosing.  

Binding kinetics differentiate medication by informing dose-response relationships. This means that physiological effectors can be appropriately counteracted by drug concentrations.  

Binding kinetics are intrinsically linked to the translation of drug action to safety, efficacy, and duration of response outcomes.  

How to Measure Binding Kinetics 

Establishing an understanding of binding kinetics is critical for developing drug molecules. To do this, equilibrium constants must be established. Koff is the dissociation constant in min-1.  Kon is the association constant in inverse minutes multiplied by inverse concentration. KD is computed from Koff/Kon. Expressed in Molar units 

Gator Bio BLI provides a simple means of measuring kinetics. Using real-time data, binding kinetics (Kon and Koff) and affinity values (KD) can be measured. Because BLI offers insight into dissociation, association rates, and affinity value it can help predict the activity of biomolecules in vivo.  

Blog

4 Key Benefits of Using BLI for Binding Kinetics Measurements

Binding kinetics make up the foundation of pharmacological theory in the modern era. Binding kinetics is a term that is concerned with the speed at which a compound binds to a target and how quickly it dissociates from it.  

BLI (bio-layer interferometry) is an optical biosensing technology used in analyzing biomolecular interactions without requiring fluorescent labeling. BLI is one of the few widely available biosensing technologies that are label-free. BLI measures macromolecular interactions by analyzing the patterns of interference from white light reflected from a biosensor tip surface. This article discusses the benefits of BLI for measuring binding kinetics. 

What Parameters of Binding Kinetics can be Determined Using BLI? 

BLI allows for real-time determination of association rate (kon) and dissociation rate (koff) in molecular interaction and as such the binding kinetics. These rates provide the opportunity to understand the overall affinity (KD) of the interaction to be calculated. This value can also be ascertained directly from the formation of concentration-dependent molecular complexes.  

Benefits of BLI 

Simplicity 

One key benefit of using BLI for binding kinetics measurements is that Kon, Koff, and KD can be measured in one rapid assay. Detection takes place in real-time without marking biological molecules. This makes the measurement quicker and simpler than processes that require multiple assays. 

Wide Ranges 

Using BLI for binding kinetics allows for a wide range of affinity values. BLI can be used for many investigations such as nanobodies, membrane proteins, antibodies, AAV, small molecules, nucleic acids, and more.  

Compatible with a Range of Samples 

BLI is compatible with both crude and purified samples. This means it can be used in a range of investigations and there is no need to swap between systems for samples. 

No Labels Required 

Label-free techniques can monitor responses in real-time, meaning kinetic information regarding dissociation and association is accessible. They can also monitor bio-molecular interactions with native binding partners without any impediment from modified analytes and therefore more accurate biochemical data is acquired.  

BLI for Binding Kinetics 

Gator Bio software provides a range of models for data analysis. Reference subtraction, processing, and data fitting can be performed quickly. Biosensors from Gator Bio combine a 1mm diameter glass rod with patented optical layers and surface chemistry specifically built at the distal end of the biosensor. To find out more about how this technology can be used in binding kinetics applications, contact the team at Gator Bio today.