Blog Taking out the trash: Guidelines for agentic single-trace BLI sensorgram processing Author: Benjamin Osborn, PhD and the Gator Bio team Overconfident BLI hit assignment impacts AI training In June 2026 the Boltz team published a candid look at how the binder-design field reports its results [1]. They noted two common points where real-time kinetics results get misread: ambiguous sensorgrams getting logged as confirmed binders, and affinities getting calculated from curve fits that don’t match the data, which makes their affinities look tighter than they are. The Boltz team proposed a nomenclature fix that can help to clarify these results: calling ambiguous cases screening hits, but reserving the term “binder” for conclusive data. In the process of validating questionable hits, they used Gator® Bio Biolayer Interferometry (BLI) instruments to invert ligand and analyte, which highlighted how kinetics can change based on how assays are designed and formatted. Experimental hit rates for AI-designed binders vary widely by project [2-4], but even in the best cases many candidates fail binding or developability criteria, so wet-lab screening remains a decisive bottleneck. And when models are trained on datasets that label candidates as binders without adequate justification, measurement errors and artifacts become overconfident training labels that bias the model’s behavior. [BACK TO TOP] A rules-based approach for BLI hit triage Existing high-throughput kinetics-analysis workflows show that preprocessing, curve fitting, residual review, and automated reporting can be standardized at scale [5-7]. This article complements that literature by addressing a narrower problem: how to interpret a first-pass BLI screen that contains only one sensorgram per candidate at a single analyte concentration. The goal is not to generate precise KD values from those traces, but to perform a rapid quality audit and assign each candidate meaningful flags and labels that inform the next steps. Instead, the goal is to sort each trace into one of the following categories: Hits: Clean candidates that pass the screening criteria Questionable hits: Candidates that pass the screen, but have one or more suspicious attributes Negative: Traces that do not achieve the hit threshold Excluded: Unusable traces that defy BLI conventions In addition to categorizing each trace, these rules will also inform downstream assays by searching for key criteria (e.g. biphasic behavior, off-rates, and bulk shifts). These in turn decide what the next experiment should be, whether it’s a classical titration for a clean hit, steady-state assays for fast-on/fast-off binders, or inverted assays for binders with unusual binding. While these pseudocode blocks are pressure-tested against real Gator instrument data, this is only meant to be a starting point for developing an AI-based BLI workflow for single-trace hit sorting. To help both humans and AIs understand the logic, there’s a section of prose explaining the rationale for these screening rules and a pseudocode section that outlines the logic clearly. Users can then query an LLM by attaching the gator data file as a .zip and using the prompt below to perform the analysis. Expand the LLM prompt ROLE You are a biolayer interferometry (BLI) data analyst. Your job is to apply a published, rule-based triage workflow to a single-trace screening dataset, exactly as written, and report the results. You are not designing new rules. INPUTS 1. The workflow article: https://www.gatorbio.com/blog/agentic-bli-sensorgram-processing/. Its B1 and B2 pseudocode blocks are the authoritative logic. The prose explains intent. 2. A .zip export from a Gator Bio instrument containing: - Assay_<n>_Channel<m>.csv files (first column = signal in nm, second column = time in s; a header row may be present; ~5 Hz) - Assay.asy and ExperimentStep.ini (step types, step times, plate map, sample names, concentrations, molecular weights) You must run real code (e.g., Python) on the data. Do not estimate from images. STEP 0: UNDERSTAND THE EXPERIMENT BEFORE ANALYZING Read Assay.asy and ExperimentStep.ini and report back, in a short list: - The step sequence and times for each assay (baseline, load, baseline, association, dissociation). If assays differ, say so. - What is loaded on the sensor (ligand) and what is in solution (analyte), with concentrations and molecular weights if listed. - Which traces are references (blank sensors, zero-analyte channels, irrelevant ligands, empty load wells). - Whether this is a single-trace screen. If it is a titration series, plan to use only the highest concentration per series. If it is anything else (for example, tandem epitope binning or a two-step association), STOP and ask how to proceed rather than forcing it into the workflow. If step times, references, or channel roles are ambiguous, ask before running. State every assumption you make. STEP 1: PREPROCESS - Apply reference subtraction as instructed by the user. If no instruction is given, propose the most appropriate reference from Step 0 and ask. - Use the raw (unsubtracted) trace for the load-step height (R_load). - Do not smooth, trim, or otherwise alter the data beyond what the pseudocode specifies. STEP 2: RUN B1, THEN B2, EXACTLY AS WRITTEN - Use the constants in the article's CONSTANTS blocks unless the user overrides one. List any override in your output. - Implement every fit fallback (seed retries, no-drift fallback, biphasic fallback with its guardrails, flat-line fallback). - Evaluate B2 classes in the stated order; the first match wins. - If a rule seems wrong for this dataset, do NOT change it silently. Run it as written, then describe the concern and its effect separately. STEP 3: OUTPUTS A. Results table, one row per trace: trace ID | sample name | disposition | class | reason | priority | next task | assoc_net (nm) | kobs | koff | t90_off (s) | dissoc_state | fails | flags B. Counts per class and disposition. C. A grid figure of all traces: - One panel per trace, shared y-axis, baseline-zeroed (subtract the mean of the last ~5 s of the pre-association baseline), showing the pre-association baseline, association, and dissociation. - Dashed horizontal line at the hit floor; light vertical lines at the association and dissociation starts. - Panel border color by outcome: HIT_CLEAN #1a9641 HIT_MODERATE_OFF #7ccf3a HIT_SLOW_ON #7b3fa0 HIT_FAST_ON_SLOW_OFF #2bb3a6 HIT_FAST_OFF #4575b4 HIT_QUESTIONABLE #f08c2e BELOW_HIT_FLOOR #bdbdbd EXCLUDE_UNUSABLE #000000 reference traces #b9b9f2 - Panel title = trace ID; add the reason for HIT_QUESTIONABLE. - Legend with class, priority, and count. D. A short plain-language summary: the hits worth following up first, anything borderline (within ~10% of a threshold), plate-level patterns (for example, failures clustered by position or run order), and any concerns about the logic or the reference choice. RULES - Every trace that clears the hit floor is a "screening hit," never a "binder." Do not report KD values; kobs and koff are triage estimates only. - BELOW_HIT_FLOOR traces are clean negatives, not failures. - Never fabricate values. If a required input (like molecular weight) is missing, skip the checks that need it and say so. - Keep sample names as given unless asked to anonymize them. [BACK TO TOP] Assumptions: The first step, not covered in this article, is running a good kinetics assay. Each of these is standard practice an experienced BLI user knows by rote, but an automated analysis strategy can only work if they’re in place. Do not mistake the brevity of this section for unimportance. Nonspecific binding has been screened out. You’ve confirmed the analyte doesn’t stick to the sensor surface on its own, independent of the ligand. The ligand is the molecule being screened; analyte is consistent. This is typical for BLI first-pass screening. Often, ligands come from cell-free media and each is screened with only one biosensor at one (often unknown) concentration. Analytes are usually at 0.1-1 µM concentration depending on the quality of starting ligands. Ligand and analyte masses are known, though they’re only used for plausibility flags. Binders are ideally monovalent, such as VHHs, scFvs, Fabs, or designed miniproteins: Full-length antibodies often show biphasic association or dissociation from avidity; the logic tolerates this but flags it. The load step is present and readable, so the logic can confirm a ligand actually captured onto each sensor. Reference subtraction is done beforehand. We recommend having a single trace with an unrelated ligand interacting with the analyte to account for the “jumps” from bulk shift as you enter and exit those wells. An unreferenced (or improperly referenced) plate may run into complications from baseline drift or bulk shift. We include baseline drift check as part of the fitting equation in case this is not performed. The workflow at a glance: This workflow is broken into two section, labeled B1 and B2 (because A will be used for steps that precede this in a later article). B1 cleans and fits the data, then decides what candidates are unusable and which ones are usable, but did not clear the hit floor. Several other flags are also fired in B1 that label the data for later sorting. B2 then takes the information that B1 has given and assigns both hit classes and priority for follow-up experiments. A flowchart of the workflow is included below in Figure 1. Figure 1: The workflow at a glance. B1 checks for whether a trace contains meaningful data, while B2 assigns priorities and next steps (colors match later figures). [BACK TO TOP] Resolving bulk shift The first thing we need to ask when examining a sensorgram is where association and dissociation actually start. The sensorgram often shows jumps going into the association and dissociation steps, caused by changes in refractive index or by binding that happened during the unread time while the sensor moved into a new well. These jumps can also appear in a reference trace, which then carries them into every trace that used that reference. We only have one trace’s information to work with to resolve this, so we measure the size of the jump on entry to the association well and on exit from it. A pure refractive index change goes up on the way in and comes back down by the same amount on the way out, so if the two jumps are opposite in direction and within 0.9 to 1.1x of each other, we treat the jump as bulk shift and start the fit from that offset. We then measure the hit against that corrected starting point, so a bulk shift can’t trigger a “hit” without any real binding. When the jumps don’t match, the direction of the mismatch tells us something. A larger entry jump suggests binding happened during the dead time before the first read, which is exactly what a very fast-on, slow-off binder looks like (JUMP_IN_EXCESS). A larger exit jump suggests signal was lost during the dead time on the way out, as with a very fast off-rate (JUMP_OUT_EXCESS). In either case, we use the smaller jump as the bulk shift estimate, since the larger one likely includes binding gained or lost while the instrument wasn’t looking. If both jumps go the same direction, it isn’t bulk shift at all; that points to a step artifact or a reference problem, and we flag it JUMP_SAME_SIGN, leaving the trace uncorrected, though we still try to see if it has useful signal. [BACK TO TOP] Preventing low-effort fitter failures Every trace gets the same fitting routine, which checks whether the binding curve behaves like 1:1 binding (or close to it). Association fit is a pseudo-first-order exponential (a global Langmuir fit needs multiple analyte concentrations; this is the single-trace application of the same model), plus a baseline drift term for ligands that slowly leave the sensor or plates without reference subtraction. Dissociation is fit as a single exponential decay. Both fits start from the bulk-corrected values described above. To keep fit failures rare, the first attempt is seeded from the data itself (e.g. the time to reach two-thirds of the rise). If it fails, we retry from several other starting points, then fall back to a model without the drift term. This catches the vast majority of BLI binders and includes samples with what appears to be a linear association phase (they just have a small koff) Some real binders don’t follow a single exponential; full-length antibodies often show a fast phase and a slow phase from avidity. When a single-exponential fit converges but exceeds the deviation limit, we try a two-phase fit with clear rules: both phases must move in the same direction, their rates must differ by at least 3x, and each must carry at least 5% of the signal and clear the noise threshold. Passing traces are flagged ASSOC_BIPHASIC or DISSOC_BIPHASIC, so the titration step won’t assume a 1:1 model. If every fit still fails, we try a flat line. If even that can’t describe the data within the deviation limit described below, the trace is erratic and excluded as ASSOC_FIT_FAILED or DISSOC_FIT_FAILED. If a flat line does fit the association, the trace is flagged ASSOC_UNRESOLVED. This will flag when a binder is so fast that its binding happens during dead time between wells. This logic provides a failsafe to protect very tight binders, but it requires us to confirm that binding happened in that dead time. The jumps from the bulk shift detection phase tell them apart: a larger entry jump (JUMP_IN_EXCESS) followed by a much smaller out-jump and accompanied by a slow or unmeasurable dissociation means the trace is kept on the strength of its off-rate alone. It may be a tight binder, but with a single trace, we can’t know enough about it to be sure. If the jumps match or there’s no slow dissociation, the trace is classified BELOW_HIT_FLOOR. If a flat line fits the dissociation, the signal simply isn’t coming off within the window, which we treat as a slow off-rate rather than a failure (covered in “Reading the dissociation phase” below). [BACK TO TOP] What is a significant deviation from a binding trace? Real binding has a characteristic shape: during association the signal rises and bends toward a plateau, and during dissociation it decays smoothly, one direction at a time. That regularity gives us something to test against. A real hit, even a noisy one, should stay within three times its noise of the fitted curve, and a trace that wanders, spikes, or refuses to settle into a binding shape is excluded. The same test applies to dissociation. We anchor this deviation limit to noise so it’s neither too harsh on noisy traces nor too loose on quiet ones. Gator Bio instruments are specified at ≤ 4 pm RMS noise at 5 Hz, and most traces run quieter than that, so we use the spec as a floor: the tightest limit is 3 × 0.004 nm = 0.012 nm. Noisier traces (for example, crude lysate) get leeway up to 0.010 nm RMS, for a limit of 0.030 nm. In the code, RESID_CAP = 3.0 is the multiplier, and the noise is clamped between 0.004 and 0.010 nm. This residual-to-noise check is our goodness-of-fit test. We avoid R², which can look acceptable on a poor fit when the binding signal is large; scaling the misfit to noise is closer in spirit to a reduced χ². Since our model is a deliberate approximation, a formal χ² analysis isn’t worth the compute, and this check handles triage well enough. Here’s one that fails. Its association passes cleanly, but then the sensor enters a well with a pipetting error. Its deviation from the decay fit is 9.5 times noise, a two-phase fit can’t rescue it, and it ends below the second baseline, earning a DISSOC_RESID fail. The cause could be an instrument fault, a bubble, the plate wall too close to the sensor tip, or insufficient well volume. The test doesn’t diagnose which; it just excludes the trace. Figure 2: example of a trace that fails the significant deviation test These fail codes also help an agent spot spatial problems on the plate: if ASSOC_RESID or DISSOC_RESID failures cluster in one region, that points to an instrument or liquid-handling issue, and those wells can be scheduled for a re-run. [BACK TO TOP] Reading the dissociation phase At screening concentrations (as high as 1 µM), an analyte can sit at 100x or more above the KD of quality binders, which can produce unusually shaped curves that are hard to tell apart from artifacts. With only one trace per candidate, we can’t rely on association for resolvable kinetics, so we turn to dissociation, usually a simple single exponential decay. Our check is the time to reach 90% dissociation, calculated from the fitted off-rate. If the signal is 90% gone within about a second, the kinetics can’t be measured. We use 1 second as a practical default for 5 Hz acquisition, since it leaves only five data points to define a decay. That corresponds to a koff of about 2.3 s⁻¹ and an average residence time of 0.43 s, beyond the instrument’s koff spec. We sort each dissociation into one of four states: FAST: 90% of the signal is gone within about a second, including signal lost before the first read (JUMP_OUT_EXCESS). NOT_RESOLVED: The off-rate is slow, but the step was too short to measure it, so the titration should run longer. This triggers when less than 5% comes off within the window, including flat lines or upward drift during dissociation. RESOLVED: everything in between, with enough decay to support the fitted off-rate. NO_SIGNAL: association never produced enough real binding to dissociate, regardless of bulk shifts. NOT_RESOLVED is where the tightest binders on a plate usually land. Their off-rate isn’t measurable yet, but B2 still treats them as “slow-off”, so they stay eligible for high priority. For an example of fast-on/fast-off kinetics, consider the two panels below. They come from the same small-molecule plate, and both pass the deviation test cleanly. The trace on the left is a hit we can characterize: association curves toward a plateau, and both rates are slow enough to read. The trace on the right rises and falls almost vertically, with a t90_off under 1 second. The binding may be real, but under these acquisition conditions, it can’t support a reliable off-rate estimate. It’s labeled FAST, and B2 sorts it into HIT_FAST_OFF at low priority. In a protein screen, an interaction that fast is usually weak; small-molecule campaigns might weigh it differently. Figure 3: Left: an example of a fast-on/fast-off interaction that is measurable. Right: a trace that is unmeasurable due to an off rate that is too fast. This trace is moved later to low priority for a steady-state screen. Both traces come from a Gator BLI small molecule experiments using SMAP (Small Molecule Analysis Probe) biosensors on a GatorPlus instrument. [BACK TO TOP] What’s a hit? We want to set a “hit floor” high enough that baseline drift won’t trip it by accident, but low enough that even a weaker ligand could trigger it. For most campaigns, a simple 0.1 nm (or higher) hit floor makes sense. It’s 10x above the maximum effective noise used by this workflow, and it’s more than any amount of baseline drift typically produced during a screening workflow’s association step (less than 180 seconds). For campaigns featuring smaller analytes or unusual ligands (like aptamers), a more permissive hit floor of 10x above noise (e.g. 10x measured noise per trace, clamped between 0.004-0.01 nm noise to control for high/low noise samples) would remain rigorous while still allowing for small analytes that may never reach 0.1 nm shifts even at saturation. An alternative to make a more stringent can be based on ligand/analyte size, and comparing overall analyte signal to load signal as a ratio. When deciding whether a trace achieves “hit” status, we look at the average of the last 5 seconds of data (measured from the bulk-corrected starting point). This lets us avoid noise on the last data point tripping the “hit” detector. Not every real binder clears the hit floor. A tight hit with a fast on-rate can reach its plateau during the dead time before the first association read, so at screening concentration its association looks flat or falls short of the floor. Two signals give it away: the entry jump is larger than the exit jump (JUMP_IN_EXCESS), because binding was added before we started reading, and the dissociation is slow or NOT_RESOLVED. When both are present, we keep the trace on the strength of its off-rate alone. B2 classifies these as HIT_FAST_ON_SLOW_OFF at medium priority; they could be the best binders on the plate, but a single trace can’t confirm it. If you’re screening at lower concentrations, this check may not be necessary. [BACK TO TOP] B1: the first batch of pseudocode There are a few pieces in the pseudocode that we didn’t talk about in the prose: Step and setup information carried with each trace: downstream analysis shouldn’t depend on the original assay.asy file always being in a particular place. HIGH_RMS, DISSOC_RISING, BASELINE_DRIFT, DISSOC_BELOW_BASELINE: quality flags that annotate a suspect trace without excluding it, giving more tools for troubleshooting. LOW_LOAD: a sensor that captured some ligand, but less than 0.1 nm or too little to clear the hit floor even at full binding (based on masses of ligand and analyte). The trace is kept, but a “non-binder” call on it may reflect poor loading rather than poor binding. NO_LOAD, by contrast, means the load was within noise and the trace is excluded. The hit floor is signed, so it also catches downward signals. Binding adds mass, so a hit is normally a rise above the floor, but unusual ligands (like aptamers) can produce real negative signals. We keep these and flag them NEGATIVE_BINDER, and B2 classifies them as HIT_QUESTIONABLE at low priority. A downward signal is more often the result of drift, a reference problem, or the analyte pulling something off the sensor, so the flag also helps trace plate positions when troubleshooting referencing errors. The values below in the CONSTANTS section are practical defaults for the assay format and acquisition rate described here, not universal BLI cutoffs. Each is tied to a stated physical or statistical rationale (above), and users are encouraged to adjust them to adapt to their assay workflow. Expand the B1 pseudocode ================================================================= B1 USABILITY + HIT FLOOR GATE 1. Is the trace physically usable? (else EXCLUDE_UNUSABLE) 2. Is there real binding, up or down? (signed hit floor) OUT: KEEP | BELOW_HIT_FLOOR | EXCLUDE_UNUSABLE All flags and fail markers start FALSE. ================================================================= --- CONSTANTS (defaults, not universal law; tune per campaign) --- NOISE_MULT = 10 # noise-relative significance HIT_FLOOR_NM = 0.10 # lower for small analytes / aptamers RMS_FLOOR = 0.004 # instrument spec at 5 Hz RMS_SIG_CAP = 0.010 # effective-noise ceiling RMS_SPEC = 0.004 # above -> HIGH_RMS flag LOAD_MIN = noise_sig # below = no ligand captured (NO_LOAD exclusion) LOAD_FLOOR_NM = 0.10 # below = too little ligand for a confident call (LOW_LOAD flag) RESID_CAP = 3.0 # deviation-from-fit cap, both phases (x rms_eff) T90_OFF_CUTOFF_S = 1.0 # 90% off faster than this = off-rate not resolvable MIN_DECAY_FRAC = 0.05 # less decay than this in-window = NOT_RESOLVED JUMP_RATIO_MIN = 0.9 # matched bulk shift: |jump_out| within 0.9-1.1x |jump_in| JUMP_RATIO_MAX = 1.1 CURVE_MIN_TAU = 1.0 # time constants elapsed for a curve to count (shared with B2) BIPHASIC_MIN_SHARE = 0.05 # each phase >= this share of total amplitude BIPHASIC_RATE_SEP = 3.0 # fast rate >= this multiple of slow rate --- NOISE -------------------------------------------------------- rms = RMS(pre-association baseline, THIS trace) rms_eff = clamp(rms, RMS_FLOOR, RMS_SIG_CAP) # see "significant deviation" prose noise_sig = NOISE_MULT x rms_eff hit_floor = HIT_FLOOR_NM --- READS (endpoints are ~5 s means, not single points) ---------- R_load = load-step signal rise R_base = mean(last ~5 s baseline, post-load) R_assoc_0 = first recorded association point R_assoc_end = mean(last ~5 s association) R_dissoc_0 = first recorded dissociation point dissoc_end = mean(last ~5 s dissociation) assoc_duration = length of association step dissoc_duration = length of dissociation step MW_ligand, MW_analyte, valency # optional; valency default 1 --- BULK SHIFT (see "Resolving bulk shift") ---------------------- jump_in = R_assoc_0 - R_base jump_out = R_dissoc_0 - R_assoc_end IF |jump_in| <= noise_sig AND |jump_out| <= noise_sig: bulk = 0 # no meaningful jump ELSE IF sign(jump_in) == sign(jump_out) AND |jump_in| > noise_sig AND |jump_out| > noise_sig: bulk = 0; flag JUMP_SAME_SIGN # can't separate bulk from binding; no correction ELSE IF |jump_out| < JUMP_RATIO_MIN x |jump_in|: bulk = -jump_out # smaller jump = bulk estimate IF jump_in > 0: flag JUMP_IN_EXCESS # upward entry = binding in dead time ELSE IF |jump_out| > JUMP_RATIO_MAX x |jump_in|: bulk = jump_in IF jump_out < 0: flag JUMP_OUT_EXCESS # downward exit = signal lost in dead time ELSE: bulk = jump_in # matched in/out = refractive index only assoc_net = R_assoc_end - (R_base + bulk) # SIGNED; feeds the hit gate --- ASSOCIATION FIT (see "Preventing low-effort fitter failures") --- # R(t) = offset + drift*t + assoc_amp*(1 - exp(-kobs*t)) # seed kobs from time-to-63%; retry seeds; fall back to no-drift assoc_fit -> offset, drift, assoc_amp, kobs, assoc_fit_ok assoc_resid = std(association - assoc_fit) / rms_eff IF assoc_fit_ok AND assoc_resid > RESID_CAP: # biphasic fallback, only when single exponential fails fit two-phase association (same drift term) IF both phases rise in the same direction AND fast rate >= BIPHASIC_RATE_SEP x slow rate AND each phase >= BIPHASIC_MIN_SHARE of total AND > noise_sig AND two-phase resid <= RESID_CAP: use two-phase fit; assoc_amp = sum of phases kobs = 1 / (time to 63% of combined rise) flag ASSOC_BIPHASIC # otherwise the single fit stands; Step 1 fails it as ASSOC_RESID IF NOT assoc_fit_ok: # flat (nothing bound, or bound in dead time) vs. erratic fit association to offset + drift only IF flat-line resid <= RESID_CAP: flag ASSOC_UNRESOLVED ELSE: assoc_fit_failed = TRUE assoc_curved = assoc_fit_ok AND (kobs x assoc_duration) >= CURVE_MIN_TAU --- DISSOCIATION FIT --------------------------------------------- # R(t) = asymptote + dissoc_amp*exp(-koff*t) # dissoc_amp may be negative, so recovery toward baseline from # either direction fits the same model dissoc_fit -> asymptote, dissoc_amp, koff, dissoc_fit_ok dissoc_resid = std(dissociation - dissoc_fit) / rms_eff IF dissoc_fit_ok AND dissoc_resid > RESID_CAP: # biphasic fallback; same guardrails as association fit two-phase decay IF guardrails pass AND two-phase resid <= RESID_CAP: use two-phase fit; koff = ln(10) / (time to 90% of combined decay) flag DISSOC_BIPHASIC # otherwise the single fit stands; Step 1 fails it as DISSOC_RESID dissoc_flat = FALSE IF NOT dissoc_fit_ok: IF flat-line resid <= RESID_CAP: dissoc_flat = TRUE # not coming off ELSE: dissoc_fit_failed = TRUE --- DISSOCIATION STATE (see "Reading the dissociation phase") -------------- dissoc_signal = |R_dissoc_0 - R_base| > noise_sig t90_off = ln(10) / koff IF koff DEFINED frac_lost = (R_dissoc_0 - dissoc_end) / (R_dissoc_0 - R_base) # fit-free IF NOT dissoc_signal AND JUMP_OUT_EXCESS: dissoc_state = FAST # lost in dead time; B2 checks association shape ELSE IF NOT dissoc_signal: dissoc_state = NO_SIGNAL # association left nothing to dissociate ELSE IF t90_off <= T90_OFF_CUTOFF_S AND |dissoc_end - R_base| <= noise_sig: dissoc_state = FAST ELSE IF dissoc_flat OR frac_lost < MIN_DECAY_FRAC: dissoc_state = NOT_RESOLVED # slow; step too short to measure it ELSE: dissoc_state = RESOLVED slow_off = dissoc_state IN {RESOLVED, NOT_RESOLVED} ================================================================= STEP 1 - USABILITY (collect ALL fails; any -> EXCLUDE_UNUSABLE) ================================================================= IF R_load < LOAD_MIN: fail NO_LOAD IF assoc_fit_failed: fail ASSOC_FIT_FAILED IF assoc_fit_ok AND assoc_resid > RESID_CAP: fail ASSOC_RESID IF dissoc_fit_failed: fail DISSOC_FIT_FAILED IF dissoc_fit_ok AND dissoc_resid > RESID_CAP: fail DISSOC_RESID IF any fail: disposition = EXCLUDE_UNUSABLE record all fails; compute flags below; STOP ================================================================= STEP 2 - HIT FLOOR (signed; direction matters) ================================================================= IF NOT ASSOC_UNRESOLVED AND assoc_net > hit_floor: disposition = KEEP ELSE IF NOT ASSOC_UNRESOLVED AND assoc_net < -hit_floor: disposition = KEEP; flag NEGATIVE_BINDER ELSE IF JUMP_IN_EXCESS AND slow_off AND (ASSOC_UNRESOLVED OR NOT assoc_curved): # buried association rescued by a real, slow off-rate # (see buried-association paragraph in "What's a hit?") disposition = KEEP ELSE: disposition = BELOW_HIT_FLOOR --- FLAGS (non-terminal; recorded regardless of disposition) ----- IF rms > RMS_SPEC: flag HIGH_RMS IF |drift x assoc_duration| > noise_sig: flag BASELINE_DRIFT IF (dissoc_end - R_base) < -noise_sig AND NEGATIVE_BINDER not set: flag DISSOC_BELOW_BASELINE IF (dissoc_end - R_dissoc_0) > noise_sig AND NEGATIVE_BINDER not set: flag DISSOC_RISING # a negative binder recovering upward is expected # low load: a "non-binder" call may reflect loading, not binding # (mass ratio is approximate in BLI; flag only) IF R_load >= LOAD_MIN AND ( R_load < LOAD_FLOOR_NM OR (MWs known AND R_load x (MW_analyte / MW_ligand) x valency < hit_floor) ): flag LOW_LOAD --- OUTPUT (self-contained record) ------------------------------- emit { disposition, fails, flags, R_load, assoc_net, bulk, jump_in, jump_out, assoc_amp, kobs, dissoc_amp, koff, t90_off, frac_lost, # kobs + koff feed B2 dissoc_state, slow_off, assoc_curved, assoc_resid, dissoc_resid, rms, rms_eff, noise_sig, hit_floor, assoc_duration, dissoc_duration, step_times, # copied in, not referenced trace_ref } # KEEP -> B2; else archived, queryable by disposition + fails + flags ================================================================= ENUMS ================================================================= DISPOSITION KEEP | BELOW_HIT_FLOOR | EXCLUDE_UNUSABLE DISSOC_STATE FAST | RESOLVED | NOT_RESOLVED | NO_SIGNAL USABILITY FAILS NO_LOAD load within noise; no ligand captured ASSOC_FIT_FAILED association erratic; no model fits, including flat line ASSOC_RESID association deviation > RESID_CAP, single and biphasic DISSOC_FIT_FAILED dissociation erratic; no model fits, including flat line DISSOC_RESID dissociation deviation > RESID_CAP, single and biphasic FLAGS (non-terminal) HIGH_RMS baseline rms above spec NEGATIVE_BINDER downward association past -hit_floor BASELINE_DRIFT fitted drift > noise_sig over association DISSOC_BELOW_BASELINE dissociation ends below pre-association baseline DISSOC_RISING signal climbs during dissociation (not negative binders); check buffer or reference JUMP_IN_EXCESS upward entry jump larger than exit; binding in dead time JUMP_OUT_EXCESS downward exit jump larger than entry; signal lost in dead time JUMP_SAME_SIGN entry and exit jumps share direction; bulk not corrected ASSOC_UNRESOLVED flat association; call rests on dissociation ASSOC_BIPHASIC two-phase association; avoid 1:1 model on titration DISSOC_BIPHASIC two-phase dissociation; avoid 1:1 model on titration LOW_LOAD load under 0.1 nm, or too low to clear hit floor at full binding [BACK TO TOP] B2: Sort the survivors Everything that reaches this point is a hit, and a secondary experiment is needed to confirm a “binder.” B1 has already set aside the unusable traces and the non-hits, so B2 only decides two things for each survivor: what kind of hit is it, and what’s next? Many screens produce more hits than can be tested, so B2 also assigns a priority. Classes are checked in this order, and the first match wins: ClassConditionNext experimentPriorityHIT_QUESTIONABLEReason: NEGATIVE, SUB_BASELINE, or KINETIC_CONFLICTOrientation flipLowHIT_FAST_OFFt90 under 1 sSteady-stateLowHIT_SLOW_ONLinear association, consistent slow offTitrationMediumHIT_FAST_ON_SLOW_OFFBuried association, JUMP_IN_EXCESS, readable off-rateTitrationMediumHIT_MODERATE_OFFCurved association, t90 between 1 and 30 sTitrationMediumHIT_CLEANCurved association, t90 ≥ 30 s or NOT_RESOLVEDTitrationHigh The low-priority classes come first on purpose, so a trace with a problem can’t be labeled a good hit by an earlier check. HIT_QUESTIONABLE traces point to the sample or the measurement rather than a well-behaved binder. KINETIC_CONFLICT is the new reason: since kobs = kon·C + koff, the off-rate can never exceed the observed association rate (we allow 1.25x for measurement error), so an association slow enough to watch paired with a dissociation too fast to read is physically inconsistent. These route to an orientation flip, which is costly enough to keep them at low priority. HIT_FAST_OFF traces lose 90% of their signal in under a second. The near-vertical, square-wave shape reflects real fast-on/fast-off binding, but usually a weak binder, so they go to a steady-state measurement at low priority. HIT_SLOW_ON traces rise in a straight line: an expected outcome at 100 nM; a very slow on-rate at 1 µM. A rise that hasn’t completed one time constant means kobs is below roughly 1 / (association time), so koff must be too. Traces that pass go to titration at medium priority; those that fail are a KINETIC_CONFLICT. Aggregation can also look like a slow, consistent linear rise, so the titration has the final word. HIT_FAST_ON_SLOW_OFF hits could be tighter than any clean hit on the plate, but because the on-rate is “buried” in dead time, a single trace can’t prove it. Placing them at medium priority rather than high is a judgment call on our part: they’re worth a titration, just not ahead of hits we can already characterize. HIT_CLEAN is the straightforward case: a curved association and a slow dissociation (a t90_off of 30 seconds or more, or NOT_RESOLVED) point to a likely good binder. HIT_MODERATE_OFF has the same characteristics, but reaches 90% dissociation in 1 to 30 seconds, which is measurable but points to a weaker binder, so it goes to titration at medium priority. A few items in the B2 pseudocode aren’t covered elsewhere in this article. RMAX_IMPLAUSIBLE flags a hit whose signal exceeds what the loaded ligand should support; it never changes the class, since BLI mass ratios are approximate, but it can assist agentic processing. plate_inputs_ok is a switch that the Rmax check reads, asking whether a whole-plate sanity check confirmed the mass and valency inputs. ASSOC_NOT_SATURATED fires when an association never reached a plateau, a note to the titration step to use a longer association or a higher concentration. These connectors are here because this stage is one piece of a larger pipeline, and it’s cheaper to leave them in place now than to retrofit them later. Expand the B2 pseudocode ================================================================= B2 CHARACTERIZE / ROUTE / PRIORITIZE [KEEP traces from B1 only] No exclusions. Every survivor is a hit; B2 assigns a class, and each class has exactly one route and one priority. Reuses B1's fits; nothing is re-measured. ================================================================= --- CONSTANTS ---------------------------------------------------- RMAX_RATIO_MAX = 1.25 # Rmax ceiling; flag only, trusted only if plate_inputs_ok KOFF_KOBS_TOL = 1.25 # koff may exceed kobs by this much before it's a conflict T90_SLOW_MIN_S = 30 # t90_off at or above this = slow off PLATEAU_TAU = 3 # ~95% of plateau reached after 3 time constants # CURVE_MIN_TAU and T90_OFF_CUTOFF_S carried from B1 --- INPUTS (from B1; nothing refit) ------------------------------ assoc_amp, kobs, koff, t90_off, dissoc_state, slow_off, assoc_curved assoc_net, assoc_duration, noise_sig, hit_floor flags carried in: NEGATIVE_BINDER, DISSOC_BELOW_BASELINE, ASSOC_UNRESOLVED, JUMP_IN_EXCESS, ASSOC_BIPHASIC, DISSOC_BIPHASIC, LOW_LOAD R_load, MW_ligand, MW_analyte, valency plate_inputs_ok = (pre-flight RMAX_INPUTS_SUSPECT did NOT fire) --- SHAPE READS -------------------------------------------------- assoc_unresolved = ASSOC_UNRESOLVED flag present # no kobs or assoc_amp exists has_curvature = assoc_curved # >= CURVE_MIN_TAU time constants # sharp = ~90% complete within the off-rate cutoff (square-wave rise) assoc_sharp = has_curvature AND (kobs x T90_OFF_CUTOFF_S) >= ln(10) curve_resolved = has_curvature AND NOT assoc_sharp # rise slow enough to watch linear_rise = NOT assoc_unresolved AND (assoc_net > hit_floor) AND NOT has_curvature assoc_buried = assoc_unresolved OR (assoc_net <= hit_floor) fast_off = dissoc_state == FAST truly_slow_off = dissoc_state == NOT_RESOLVED OR (dissoc_state == RESOLVED AND t90_off >= T90_SLOW_MIN_S) not_plateaued = NOT assoc_unresolved AND (kobs x assoc_duration) < PLATEAU_TAU neg_binder = NEGATIVE_BINDER flag present sub_baseline = DISSOC_BELOW_BASELINE flag present --- KINETIC CONSISTENCY (see kinetic-conflict prose) ------------- # kobs = kon*C + koff, so koff can never exceed kobs. # Curved rise: compare koff to the fitted kobs. # Linear rise: fewer than CURVE_MIN_TAU time constants elapsed, so # kobs < CURVE_MIN_TAU / assoc_duration; koff must sit under that bound. # Only RESOLVED off-rates are compared; NOT_RESOLVED koff is an extrapolation. kobs_max = CURVE_MIN_TAU / assoc_duration IF linear_rise koff_exceeds_kobs = has_curvature AND dissoc_state == RESOLVED AND koff > KOFF_KOBS_TOL x kobs koff_exceeds_max = linear_rise AND dissoc_state == RESOLVED AND koff > KOFF_KOBS_TOL x kobs_max kinetic_conflict = koff_exceeds_kobs OR koff_exceeds_max OR ((curve_resolved OR linear_rise) AND dissoc_state IN {FAST, NO_SIGNAL}) # per-candidate Rmax ceiling; amplitude only trusted on a curved fit Rmax_pred = R_load x (MW_analyte / MW_ligand) x valency # valency default 1 rmax_ratio = assoc_amp / Rmax_pred IF has_curvature rmax_breach = plate_inputs_ok AND has_curvature AND (rmax_ratio > RMAX_RATIO_MAX) --- CLASS (low priority first; pull problems out early) ---------- IF neg_binder: class = HIT_QUESTIONABLE; reason = NEGATIVE ELSE IF sub_baseline: class = HIT_QUESTIONABLE; reason = SUB_BASELINE ELSE IF kinetic_conflict: class = HIT_QUESTIONABLE; reason = KINETIC_CONFLICT ELSE IF fast_off: class = HIT_FAST_OFF # real fast-on/fast-off: sharp rise, fast decay ELSE IF linear_rise: class = HIT_SLOW_ON # passed the kobs_max check above ELSE IF assoc_buried AND slow_off AND JUMP_IN_EXCESS: class = HIT_FAST_ON_SLOW_OFF # rescued in B1 on dissociation alone ELSE IF truly_slow_off: class = HIT_CLEAN ELSE: class = HIT_MODERATE_OFF # measurable, but comes off within ~30 s --- ROUTE + PRIORITY (fixed per class) --------------------------- HIT_CLEAN -> TITRATION HIGH HIT_MODERATE_OFF -> TITRATION MEDIUM HIT_SLOW_ON -> TITRATION MEDIUM HIT_FAST_ON_SLOW_OFF -> TITRATION MEDIUM HIT_FAST_OFF -> STEADY_STATE LOW HIT_QUESTIONABLE -> ORIENTATION_FLIP LOW --- FLAGS (appended; never change class) ------------------------- IF rmax_breach: flag RMAX_IMPLAUSIBLE IF not_plateaued: flag ASSOC_NOT_SATURATED # titration: longer assoc / higher conc --- OUTPUT ------------------------------------------------------- emit { class, reason, route, priority, has_curvature, assoc_sharp, curve_resolved, linear_rise, assoc_buried, dissoc_state, truly_slow_off, kobs, kobs_max, koff, t90_off, assoc_amp, rmax_ratio, flags += {RMAX_IMPLAUSIBLE?, ASSOC_NOT_SATURATED?} } ================================================================= ENUMS ================================================================= CLASS HIT_CLEAN curved, slow off (t90 >= 30 s or NOT_RESOLVED) -> titration, HIGH HIT_MODERATE_OFF curved, t90 between 1 and 30 s -> titration, MEDIUM HIT_SLOW_ON linear rise, koff within kobs_max bound -> titration, MEDIUM HIT_FAST_ON_SLOW_OFF buried association, readable off-rate -> titration, MEDIUM HIT_FAST_OFF t90 under 1 s -> steady-state, LOW HIT_QUESTIONABLE see REASON -> flip, LOW REASON (HIT_QUESTIONABLE only) NEGATIVE downward association SUB_BASELINE dissociation ends below baseline KINETIC_CONFLICT off-rate contradicts association (fitted kobs, or kobs_max for linear rises) ROUTE TITRATION | STEADY_STATE | ORIENTATION_FLIP PRIORITY HIGH | MEDIUM | LOW FLAGS (appended in B2) RMAX_IMPLAUSIBLE fitted ceiling > RMAX_RATIO_MAX x predicted ASSOC_NOT_SATURATED association didn't plateau; longer assoc / higher conc on titration [BACK TO TOP] Applying the sorter Figure 4 is an experimental set with 48 different antibody ligands against the same analyte run on a Gator Plus instrument. I ran this data through B1 then B2 and asked the LLM to apply a colored border based on the outcome. Most read as non-binders (grey), exactly what a yes/no screen should produce. The survivors sort cleanly into the HIT_CLEAN (dark green) and HIT_MODERATE_OFF (light green) classes. We also see a few HIT_SLOW_ON (purple) on this plate, which makes sense given that the analyte was run at 100 nM. These may be slower-on binders, but their dissociation is slow and consistent, so they’re worth a titration. A messier plate, or one with faster kinetics, might also populate the steady-state and flip classes too. A keen eye might point out a sample like Assay 4, channel 5: why is it still getting classified as a high priority binder when it just barely clears the “hit” bar? This is because we’re using a uniform 0.1 nm hit threshold, and because its t90_off is about 60 seconds, past the 30 second cutoff. In all respects, it meets our HIT_CLEAN parameters and the pseudocode is behaving as intended. However, going into an assay like this, I might set a slightly higher hit threshold given the size of the analyte. That said, be wary of setting a dynamic threshold that relies on the association height of the other hits on the plate (e.g. cutting all hits below 25% of the median of all hits). Simple logical rules of this nature can backfire and omit perfectly good binders when all hits are around the same intensity, or when there are few hits on the plate. Figure 4: A 48-trace BLI high throughput single-trace screen featuring 11 high priority hits. No referencing was used for these samples. Figure 5 is a 60-antibody screen against a 100 nM analyte concentration, with 300-second association and dissociation steps and four empty sensors averaged as the reference. Here, the sorter prioritizes each trace as expected. Some traces that look like fast-offs, such as A4Ch1, reach 90% dissociation at 34 seconds, just past our 30-second cutoff, so they’re classified as high priority alongside the other binders in green. Eight candidates survive, all high priority, and the remaining 52 fall below the hit floor as clean negatives. Assay 3 Channel 1 and Assay 8 Channel 4 both show the slowest off-rates (t90 of 250 and 350 seconds, respectively), making them strong candidates for follow-up. Figure 5: a 60-antibody high throughput single-trace screen with 8 high priority hits identified. Samples A8Ch5-A8Ch8 were averaged together as reference traces. To get an example of clear biphasic behavior, we ran it against a dataset that features Fc receptors as ligand and full-length IgG as analyte. While the assay was originally a titration experiment (each assay its own titration with one Ab tested in 1-3 and another in 4-6 against three different Fc receptors), the instructions to the LLM included a statement to use only channel 1 (the top concentration of each titration series: 30 µM for Assays 1, 2, 4, and 5; 10 µM for Assays 3 and 6) and to use Channel 8 from each assay used as reference. As we explain in the assumptions section, most screening campaigns won’t see this kind of fast-on, biphasic binding. However, the logic correctly assigned Assays 1 and 4 as HIT_CLEAN at high priority and 2 and 5 as HIT_MODERATE_OFF based on the dissociation rate. Assay 6 falls below the hit floor, a clear negative. Assay 3 is the interesting one. It clears the hit floor, but its fitted off-rate exceeds its observed association rate, which points to a problem with the sample or the measurement. B2 classifies it as HIT_QUESTIONABLE (KINETIC_CONFLICT) at low priority, since the check can’t tell which. Figure 6: two antibodies tested against Fc receptors. Only top concentrations are considered, with reference channels applied per assay (not shown). Assays 1-3 are Ab1 against three different Fc receptors; Assays 4-6 are Ab2. This was an added test to confirm functional sorting logic in B1/B2 for biphasic binders. Finally, we have an example of some uglier data, Figure 7. This is a kinetic screen of 17 antibodies captured on the sensor, run against a single target protein. We used only the top concentration (100 nM) and subtracted a sensor that captured no antibody as the reference. Four antibodies (Ab13, Ab15, Ab16, and Ab17) are HIT_CLEAN: they curve toward a plateau and barely dissociate over five minutes. Five more (Ab1 through Ab5) clear the hit floor but rise in a straight line without curving, and none of them dissociate measurably in the window (NOT_RESOLVED). That pairing is self-consistent, so B2 classifies them as HIT_SLOW_ON at medium priority. One lesson from this dataset was how much the reference matters: the first choice of reference had a small step at the start of dissociation, which the fitter read as a fast off-rate and flagged a kinetic conflict on two good binders. A cleaner reference (A5Ch7, an irrelevant antibody sensor that captured almost nothing), made both conflicts disappear. Figure 7: Example dataset with 17 different binders. This result required using a different trace than what is labeled as “blank” (A5Ch5) to obtain a usable reference. After using A5Ch7 for reference, all traces resolved cleanly. [BACK TO TOP] Some caveats: Automated analysis rules and flags can cut down the workload for a reviewer, allowing them to quickly train an AI model with this starting foundation. However, this workflow cannot replace a trained kinetics expert. Some things to bear in mind (which we’ll cover in more depth in a follow-up article): Referencing: A reference sensor loaded with an irrelevant ligand usually matches the sample surface better than a blank biosensor. A loaded sensor seeing zero analyte is also not a sufficient reference. It may account for baseline drift, but it doesn’t see the bulk shift jumps as you enter the high analyte concentration. If a consistent set of spikes are seen across candidates, try double-referencing using a ligand that showed no binding and see if that improves your fits. Rebinding: on a densely loaded sensor, analyte that comes off can rebind a neighboring ligand before it escapes. Both slow the apparent off-rate and make a trace appear to have a slower or biphasic off rate. This means that the koff from the initial screen may not align with the final koff of a titration performed at a concentration closer to the KD. DISSOC_RISING: flags traces whose signal climbs during dissociation, which can mask a real off-rate. If it appears across many traces that share the same analyte or buffer, the buffer itself is the likely cause (for example, a bad lot of BSA), and the fix is a fresh buffer or a better-matched reference rather than re-running the candidates. [BACK TO TOP] A step toward automated rigor None of the individual rules here are unique. A bench scientist makes most of these calls by eye without thinking about them. However, at screening scale nobody has time to make those calls by hand across hundreds of candidates, so the work here was less about inventing new science than about writing down what a careful analyst already knows. This piece is in the middle of the workflow, focusing on the steps of garbage collection and hit sorting. Future articles will examine: Pre-flight quality checks for BLI data and how to set up a high throughput screen with the greatest chances of success What a follow-up assay looks like for each condition (steady-state vs kinetics titration) and how to know you’ve generated good data. In the meantime, feed the logic above into your favorite LLM, pull one of your own screens (or download this one, corresponding with the 48 trace panel above), and see where it agrees with you and where it doesn’t. We’d love to hear how it performed: reply to our post on LinkedIn and let us know. To hear more about how Gator Bio is powering AI research and hit screening, check out our recent recordings from Closing the Loop SF 2026, a gathering on the AI-native lab. [BACK TO TOP] Works Cited [1] Boltz team. De novo binder design model evaluation. Boltz; 2026 [cited 2026 Oct 5]. Available from: https://boltz.com/experimental-protocol [2] Erasmus, M.F., Bedinger, D., Hopkins, E. et al. A blinded, prospective benchmark of in silico antibody discovery anchored to experimental affinity and developability. Nat Biotechnol (2026). https://doi.org/10.1038/s41587-026-03238-6 [3] Bennett NR, Coventry B, Goreshnik I, et al. Improving de novo protein binder design with deep learning. Nature Communications. 2023;14:2625. doi:10.1038/s41467-023-38328-5. [4] Dreyer FA, Schneider C, Kovaltsuk A, et al. Computational design of therapeutic antibodies with improved developability: efficient traversal of binder landscapes and rescue of escape mutations. mAbs. 2025;17(1):2511220. doi:10.1080/19420862.2025.2511220. [5] Srinivasan B, Flórez Weidinger JD, Zhai X, Lemercier G, Ikeda T, Brewer M, Zhang B, Heyse S, Wingfield J, Steigele S. High-throughput mechanistic screening of non-equilibrium inhibitors by a fully automated data analysis pipeline in early drug-discovery. SLAS Discov. 2022 Dec;27(8):460-470. doi:10.1016/j.slasd.2022.09.003. Epub 2022 Sep 23. PMID: 36156314. [6] Dahl G, Steigele S, Hillertz P, Tigerström A, Egnéus A, Mehrle A, Ginkel M, Edfeldt F, Holdgate G, O’Connell N, Kappler B, Brodte A, Rawlins PB, Davies G, Westberg EL, Folmer RH, Heyse S. Unified Software Solution for Efficient SPR Data Analysis in Drug Research. SLAS Discov. 2017 Feb;22(2):203-209. doi:10.1177/1087057116675316. Epub 2016 Oct 28. PMID: 27789754; PMCID: PMC5302086. [7] Li K, Huntwork RHC, Horn GQ, Alam SM, Tomaras GD, Dennison SM. TitrationAnalysis: a tool for high throughput binding kinetics data analysis for multiple label-free platforms. Gates Open Res. 2024 Jun 28;7:107. doi:10.12688/gatesopenres.14743.2. PMID: 38009106; PMCID: PMC10667272.